PN 4237523A
3-17
OPERATION PRINCIPLES
PARAMETERS AND THEIR DERIVATION
3
Absolute Numbers
The absolute numbers of leukocytes in each category are computed from the WBC count and
the differential percentage parameters.
Reticulocyte (Retic) Parameters
A graphic display of cell populations in a scatterplot form is available for reporting. A flag
indicates the sample needs further investigation. Results then should be interpreted by a
physician or other qualified, medical professional.
Reticulocyte Percent (RET%)
This is the number of reticulocytes per 100 RBCs. This parameter is directly measured and
reported as %, a percentage of RBCs.
Reticulocyte Absolute Number (RET#)
This is the absolute number of reticulocytes. It is calculated from the reticulocyte percent
(RET%) and red cell (RBC) count. It is expressed as the number of reticulocytes per liter and
reported in the same units selected by the user for RBC.
On the HmX Hematology Analyzer, you can enter the red cell count manually, via data base
query, if you want the absolute number result. For more information, see the heading,
Cycling Retic Samples, in the Operator's Guide.
NE# 10
3
cells/
µ
L
(
)
NE%
100
-------------
WBC count
×
=
LY# 10
3
cells/
µ
L
(
)
LY%
100
-------------
WBC count
×
=
MO# 10
3
cells/
µ
L
(
)
MO%
100
--------------
WBC count
×
=
EO# 10
3
cells/
µ
L
(
)
EO%
100
-------------
WBC count
×
=
BA# 10
3
cells/
µ
L
(
)
BA%
100
-------------
WBC count
×
=
RET#
RET%
RBC Count
×
100
--------------------------------------------------
=
Summary of Contents for HmX Hematology Analyzer
Page 4: ...PN 4237523A iv REVISION STATUS ...
Page 14: ...PN 4237523A xiv CONTENTS ...
Page 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Page 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Page 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Page 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Page 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Page 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Page 98: ......
Page 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Page 124: ...PN 4237523A 14 INDEX INDEX ...
Page 126: ......