PN 4237523A
4-8
SPECIFICATIONS/CHARACTERISTICS
PERFORMANCE CHARACTERISTICS - HmX HEMATOLOGY ANALYZER
Limits over the clinical range for a minimum of 50 specimens from a general hospital
population of no more than 30% specimens with abnormally elevated Reticulocyte values
(Retic >4%) are:
Limits over the clinical range for a minimum of 50 specimens with the following
characteristics are:
r
Greater than 50% abnormally elevated Retic values (Retic >4%).
r
No greater than 5 of 50 specimens have Retic values >20%.
* Both requirements must be met.
Reportable Range
Table 4.7 shows the reportable range for the HmX Hematology Analyzer with Reticulocyte
method.
Operating Range
The operating range is 0 - 100 percent reticulocytes.
Flagging
Samples above or below the reportable range are flagged accordingly. Instrument conditions
that may interfere with the identification of reticulocytes are also flagged. Refer to the
Parameter Codes table in the Operator's Guide.
Limitations
Prepare samples at ambient temperatures between 16 and 30°C (61 and 86°F). Use air
displacement pipettes to pipet whole blood and to transfer the blood/stain mixture. Repeated
aspiration of Reagent A can cause pipette discoloration. Also see Heading 4.8, INTERFERING
SUBSTANCES.
4.5
PERFORMANCE CHARACTERISTICS - HmX HEMATOLOGY ANALYZER
Precision of the CBC Parameters
The tests described in this section were performed on the WBC differential and CBC
subsystems as separate, noninteractive entities. This test condition is dictated, in part, by the
need to use separately validated reference methods for evaluation of instrument accuracy and
Parameter
Mean Difference*
SD of Difference*
RET%
±1.0
≤
1.5
Parameter %
Mean Difference*
SD of Difference*
0.00 - 30.00
±1.5
≤
3.0
Table 4.7 Reportable Range
Parameter
Range
Mean Difference
Reticulocyte %
0.2 - 30.0
±0.3
Summary of Contents for HmX Hematology Analyzer
Page 4: ...PN 4237523A iv REVISION STATUS ...
Page 14: ...PN 4237523A xiv CONTENTS ...
Page 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Page 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Page 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Page 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Page 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Page 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Page 98: ......
Page 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Page 124: ...PN 4237523A 14 INDEX INDEX ...
Page 126: ......