PN 4237523A
3-8
OPERATION PRINCIPLES
SAMPLE FLOW
8.
The BSV rotates back. The Erythrolyse II (PAK LYSE) reagent pumps send the diff lytic
reagent to pick up and deliver the diff portion to the mixing chamber, rupturing red cell
membranes and dissolving cell debris in the process.
The StabiLyse (PAK PRESERVE) reagent pump dispenses diff leukocyte preservative. It
enters the mixing chamber to stabilize the leukocyte subpopulations. See Figure 3.8.
Figure 3.8 Erythrolyse II (PAK LYSE) and StabiLyse (PAK PRESERVE) Reagent Pumps
9.
The vacuum isolators supply vacuum to the baths to pull the sample through the
apertures for the cell counts. The unit counts and sizes RBCs and Plts at the RBC
aperture and WBCs at the WBC aperture.
It measures hemoglobin photometrically through the WBC bath. See Figure 3.9.
Figure 3.9 Aperture Baths and Vacuum Isolators
Mixes
Erythrolyse II
reagent pumps
StabiLyse
reagent
pump
Vacuum
isolators
Hgb
lamp
Hgb
photometer
card
WBC
aperture
bath
Waste
chamber
RBC
aperture
bath
Summary of Contents for HmX Hematology Analyzer
Page 4: ...PN 4237523A iv REVISION STATUS ...
Page 14: ...PN 4237523A xiv CONTENTS ...
Page 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Page 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Page 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Page 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Page 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Page 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Page 98: ......
Page 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Page 124: ...PN 4237523A 14 INDEX INDEX ...
Page 126: ......