PN 4237523A
7-3
BAR-CODE SPECIFICATIONS
BAR-CODE LABELS
7
Bar-code labels used in the system must be grade C or better, according to American National
Standards Institute's guidelines, X3.182-1990.
Label Specifications
See Figure 7.1 for bar-code label specifications.
Figure 7.1 Bar-Code Label Specifications
IMPORTANT
Use of bar codes is an extremely accurate and effective method of data capture. Certain
features, such as checksum digits, maximize accuracy in reading Codabar, NW7, Code 39, and Interleaved
2-of-5 labels. In one study, the use of checksum digits detected 97% of misread errors.
Use checksums to provide protection against occasional misread errors caused by problems such as
damaged or misapplied labels. If you must use bar codes without checksums, it is recommended that you
verify each bar-code reading to assure correct patient identification.
Leading edge of label
0.055"
0.005"
0.100 + 0.030"
0.062"
630-370
Label length
10X
min.
All subsequent bar lines parallel to first bar
line within 0.001"
Human-readable code
Placement
indicator
*
Maximum thickness of label +
adhesive = 0.006 ±0.003"
0.045 +0.045"
0.040 +
0.030"
First bar line parallel to leading edge within 0.002"
Trailing quiet zone
Code area
(Start character, data digits,
checksum, stop character.)
X= narrow element width
leading quiet zone
10X
min.
0.300"
min.
Label width
maximum 1.1"
minimum 0.40"
(You must leave at least 1/8"
so you can see tube contents.)
Summary of Contents for HmX Hematology Analyzer
Page 4: ...PN 4237523A iv REVISION STATUS ...
Page 14: ...PN 4237523A xiv CONTENTS ...
Page 18: ...PN 4237523A xviii INTRODUCTION COMPUTER PROGRAM STATEMENT ...
Page 28: ...PN 4237523A 1 10 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 34: ...PN 4237523A 2 6 INSTALLATION INTERUNIT CONNECTIONS ...
Page 52: ...PN 4237523A 3 18 OPERATION PRINCIPLES PARAMETERS AND THEIR DERIVATION ...
Page 70: ...PN 4237523A 4 18 SPECIFICATIONS CHARACTERISTICS INTERFERING SUBSTANCES ...
Page 78: ...PN 4237523A 6 4 REPORTING OPTIONS GRAPHIC FORMAT ...
Page 84: ...PN 4237523A 7 6 BAR CODE SPECIFICATIONS BAR CODE DECODER ...
Page 98: ......
Page 110: ...PN 4237523A 8 GLOSSARY GLOSSARY ...
Page 124: ...PN 4237523A 14 INDEX INDEX ...
Page 126: ......