PN 4237523A
3-5
OPERATION PRINCIPLES
SAMPLE FLOW
3
r
After aspiration, the needle is retracted.
r
The sample tube is ejected from the carousel.
r
Sample analysis occurs.
r
The Ready light is green to indicate another sample can be processed.
3.4
SAMPLE FLOW
Normal Sample Flow
1.
The appropriate aspiration pump pulls the sample into the BSV (Figure 3.5):
r
125 µL in the open vial mode
r
185 µL in the closed vial mode.
Figure 3.5 BSV
ASPIRATION
28 µL diff loop
28 µL WBC loop
1.6 µL RBC