Fig. 4-28
11.
When the sample fails, the display will show the test results (maximum load and
resistance) and the system will emit an audible alarm (5 beeps) to inform the
operator; the trend indicator will also show the symbol
*****
. In case the full scale
of the system was reached before the failure of the sample, the system will only emit
the audible alarm;
Fig. 4-29
12.
Turn OFF the pump motor by pressing the relevant switch, then turn lever B slowly
back and lever A down to discharge the oil pressure;
Fig. 4-30
A
B