The automatic test procedure, once the specimen has been positioned and centered and
the machine switched on is:
•
Set test parameters, including load rate, on the touch screen (can be avoided by
default for repetitive tests);
•
Press the start button on the touch screen;
•
Move the oil delivery valve upwards;
•
The machine automatically starts the rapid platen approach, switches to the test
speed once the specimen comes into contact with the upper platen and, finally,
releases the pressure upon specimen failure;
•
Save the test including data results and load/time graph. The test execution
conforming to standards can be easily proved;
•
Move the oil delivery valve downwards.
Main features are:
•
Multipurpose machine ideal for Central laboratories, for testing Cement, Mortars,
Resins, Refractory, Lightweight concrete, Soil-cement specimens etc;
•
digital control system with wide high-contrast display 4x20 characters and 6 keys
membrane keyboard;
•
real time display of load and stress and applied load rate by symbols for easy
adjustment;
•
accuracy Class 1 (EN) and Class A (ASTM) starting from 10% of full scale;
•
special calibration starting from few kilonewtons available on request;
•
optional control of a second frame for single chambers testers. Active frame selection
via manual lever;
•
optional internal graphic printer;
•
universal compression testers, in practice, suitable to house various accessories for
testing cement, mortar, resins and other materials;
•
LAN port for PC connection and data transmission during test execution;
•
Data storage on removable USB pen drive.