106
13.
Enter the first load step (typically 0) on the left column of the first row on the display
(use the
keys to select the digits to change and the
keys to set the figure), then press the
key to access the figure on
the right column on the display and enter the corresponding bit reading previously
recorded;
14.
Press again the
key to access the second row, left column, enter the load
step, press
to access the right column row and enter the corresponding bit
reading previously noted for this step;
15.
Once all rows on the first page are completed, pressing the
key brings to
the second page with rows from 4 to 6;
16.
Proceed in this way until entering all the load steps and bit readings previously
recorded. At the end press the
key to complete the calibration procedure,
the system will ask for a confirmation;
Fig. 5-27
17.
Press the
key to accept the new calibration or the
key to exit
without altering the existing calibration. The system will return to the
MAIN
MENU
;
18.
Repeat the calibration verification (see chapter 5.2.5.1) to make sure the system is
now behaving linearly within the required load range and tolerance.