1.1 Introduction

The CONMED 1000SES

TM

Smoke Evacuation

System is designed to remove smoke and noxious
odors produced by surgical smoke during laser
and electrosurgical procedures. 

The CONMED 1000SES

TM

has been designed

with a high suction, high flow rate centrifugal
action pump. The ultra-quiet motor is used to
draw the surgical smoke from the surgical site,
through the vacuum tubing and into the
CONMED 1000SES

TM

filter where the surgical

smoke is processed by a series of filters. A single
disposable filter is used to simplify the installation
and removal during filter changes. The filter is
completely enclosed to protect the healthcare per-
sonnel from potential contamination during filter
changes. One CONMED 1000SES

TM

Filter con-

tains four different filter stages to decontaminate
the smoke plume. 

The first stage filtration is a prefilter whose func-
tion is to trap and remove gross particulate and
casual fluid. 

The second stage filtration is a VLSI (Very Large
Scale Integrated) grade ULPA (Ultra Low
Penetration Air) filter whose high-tech patented
design captures particulates and micro-organisms
down to 0.012 µm at an efficiency of
99.99995%. Each VLSI grade ULPA filter is pre-
tested and certified before use.

The third stage filtration uses the highest grade
virgin coal grade activated carbon for the removal
and adsorption of odors and toxic gases produced
by cauterization of tissue. These harmful gases
may constitute a health hazard to healthcare pro-
fessionals who are subjected to prolonged expo-
sure. The activated carbon used in the CONMED
1000SES

TM 

preferentially removes toxic organic

gases rather than water vapor and provides opti-
mal odor removal.

The fourth stage filtration is an expanded foam
and paper media used to trap activated carbon
fines from migrating out of the filter. The first,
second and fourth stages of the CONMED
1000SES

TM

Filter have been impregnated with a

chemical agent. The chemical agent has been
applied with a patented process that prevents the
agent from being washed or blown off the filter
media. 

The electronic controls on the face panel of the
CONMED 1000SES

TM

have been designed “user

friendly” and facilitate unit set up and operation.
Please refer to Section 2.0 for Operating
Instructions.

1.2 Inspection

The CONMED 1000SES

TM

Smoke Evacuator has

been thoroughly tested and inspected before ship-
ment from the factory. Please check the unit
before using it to insure that no damage has
occurred in transit. If damage is evident, please
contact CONMED Customer Service at one of
the numbers listed on the inside front cover of
this manual.

The unit as shipped also includes an AC Power
Cord for connection to a standard outlet.

A list of available accessories for the CONMED
1000SES

TM

is presented in Section 4.3. Please

contact CONMED Customer Service at the num-
bers listed on the inside front cover of this manu-
al to order accessories.  

1.3 Cautions and Warnings

Please note that all Cautions and Warnings should
be read and understood before any use of this
equipment.

This smoke evacuation system should only be
operated by personnel who have been properly
trained in its operation.

System Description

Section 1.0

1-1

60-6852-ENG (-).qxd  2/1/99  12:40 PM  Page 1-1

Summary of Contents for 1000 SES

Page 1: ...Service Manual English Version Smoke Evacuation System...

Page 2: ...han intended 1999 CONMED Corporation Mailing Address 310 Broad Street Utica New York 13501 U S A Shipping Address 7211 S Eagle St Englewood Colorado 80112 U S A For Technical Services or Return Author...

Page 3: ...spection 3 1 3 4 Troubleshooting the System 3 2 4 0 CUSTOMER SERVICE 4 1 4 1 Equipment Return 4 1 4 2 Ordering Information 4 1 4 3 Available Accessories 4 1 5 0 THEORY OF OPERATION 5 1 5 1 Filtration...

Page 4: ...1 Control Panel 2 1 2 Back View 2 4 3 Front View 2 4 4 Air Flow Diagram 5 2 5 System Block Diagram 5 3 6 Exploded View 6 5 7 Insulation Assembly 6 6 8 Electrical Schematic 6 7 Figure Title Page...

Page 5: ...subjected to prolonged expo sure The activated carbon used in the CONMED 1000SESTM preferentially removes toxic organic gases rather than water vapor and provides opti mal odor removal The fourth stag...

Page 6: ...CHNICAL PERSONNEL There are no user serviceable components in the CONMED 1000SESTM Smoke Evacuation System Refer service to qualified service per sonnel Use only with the AC cord provided and always p...

Page 7: ...WN arrow button decreas es suction The suction control should be set at the lowest setting to completely remove the surgi cal smoke from the operative site Each time either arrow button is depressed t...

Page 8: ...le signs of structural damage 2 Insert the CONMED 1000SESTM Filter into the round opening in the front of the smoke evac uation system Be sure that the filter is seated completely against the rear of...

Page 9: ...f surgical smoke When two indicators are illuminated on the suction con trol scale the suction control setting is the value between the two illuminated percentages 8 Deactivate the unit by a Pressing...

Page 10: ...Breakers Remote Switch Jack Cautions Product Nameplate I 0 IPX1 Control Panel Filter Inlet Filter CONMED 1000SES Handle Cooling Vents Power Switch Remote Switch Jack Circuit Breakers Cautions Power Co...

Page 11: ...ng it with a cloth that has been dampened not dripping with a disinfectant or a mild soap or detergent solution such as Windex or Formula 409 Windex is a registered trademark of the Drackett Products...

Page 12: ...Tubing 4 Call BioMed or CONMED Tech Services at the numbers listed on the inside front cover of this manual 1 Check power outlet and connection to rear panel of machine 2 Reset circuit breaker on bac...

Page 13: ...001 Remote Switch Adapter 10 Cord Cat No 60 6886 002 Remote Switch Adhesive Pad Bag of 5 Cat No 60 6887 001 CONMED Disposable Smoke Evacuation Handpiece Case of 5 Cat No 60 6800 001 Hand Held Disposa...

Page 14: ...4 2 This page intentionally left blank...

Page 15: ...hose healthcare workers who breathe the gases For this reason the highest quality virgin coal grade activated carbon is used This activated car bon preferentially removes toxic organic gases rather th...

Page 16: ...lled by a triac on the con troller PWB and runs on mains power This motor blower is rated at 115 CFM and a static suction of 97 H2O Replacement brushes for this AC motor are available from CONMED Tech...

Page 17: ...5 3 Figure 5 System Block Diagram...

Page 18: ...5 4 This page intentionally left blank...

Page 19: ...mine the entire system for physical damage Contact BioMed or CONMED for service if damaged Yearly Inspections Perform the chassis ground integrity test per Section 6 3 Check the Service light on the m...

Page 20: ...uum hose clogged 3 Check or replace hose 4 Motor Blower obstructed 4 Contact CONMED Technical Services 2 CONMED 1000 SES 1 Not plugged into electrical 1 Check power outlet or check to does not functio...

Page 21: ...If necessary carefully remove the insulation from the base If any insulation is removed it may not be reinstalled because the adhesive may not func tion properly Newer models do not have adhe sive and...

Page 22: ...luminate after 500 hours of use This is an indication to service the motor brushes Upon inspection if you find the carbon part of the motor brush is long or less replace the brushes Replacement brushe...

Page 23: ...6 5 Figure 6 Exploded View...

Page 24: ...6 6 Figure 7 Insulation Assembly...

Page 25: ...6 7 Figure 8 Electrical Schematic...

Page 26: ...h Jack Wire Harness 1 9 CHP C1X0 8 Microprocessor IC REV 1 1 10 WIRE ASSBLY Wire Assembly w E Terms 1 11 GASKET LED LED Gasket 2 12 MEMBRANEBLNK Membrane Switch Panel Blank 1 13 MEMBRANE CON CONMED Me...

Reviews: