3-2
POTENTIAL CAUSE
1. Filter not seated completely.
2. Filter clogged.
3. Vacuum hose clogged.
4. Motor/Blower obstructed.
1. Not plugged into electrical
outlet.
2. Circuit breaker(s) tripped.
3. Footswitch plugged in.
4. Electronic system failure.
5. Main power switch in OFF
position.
CORRECTIVE ACTION
1. Re-install CONMED
1000SES
TM
Filter, press firmly
into place.
2. Replace filter.
3. Check or replace Smoke
Evacuation Tubing.
4. Call BioMed or CONMED
Tech Services at the numbers
listed on the inside front cover
of this manual.
1. Check power outlet and
connection to rear panel of
machine.
2. Reset circuit breaker on back
of unit.
3. Unplug footswitch.
4. Call BioMed or CONMED
Tech Services at the numbers
listed on the inside front cover
of this manual.
5. Check main power switch
on rear of machine to ensure
that it is in the ON (
I
) posi-
tion.
3.4 Troubleshooting the
System
PROBLEM
1. Smoke Evacuation System is
on but suction is minimal or
none.
2. Smoke Evacuation System
does not function even though
on button is depressed.
60-6852-ENG (-).qxd 2/1/99 12:40 PM Page 3-2