ConMed 1000 SES Service Manual Download Page 9

2-3

2. Attach unit power cord to the receptacle on the
rear panel of the machine. Plug the three prong
power connector into an appropriate power out-
let.

3. Optional:  Attach either Remote Activator
Switch or Footswitch by inserting the male
footswitch plug into REMOTE SWITCH recep-
tacle on rear of machine.

4. Attach the CONMED Smoke Evacuation
Pencil or tubing set to the inlet of the CONMED
1000SES

TM

Filter.  Ensure that the evacuation

tubing is fully seated on the inlet of the filter.

5. Turn on unit main power by placing the power
switch on the rear panel of the machine in the
“ON” position.

6. Activate the power unit by:

a. Pressing the ON/OFF switch on the

membrane panel,

b. Depressing and releasing the footswitch (if 

connected), or

c. Activating either CUT or COAG on the

CONMED Smoke Evacuation Pencil (if
Remote Activator is connected).

7. Adjust the suction level to the desired setting
by pressing the UP and DOWN arrow keys while
the unit is activated.  Noise created by the smoke
evacuation power unit may be minimized by
selecting the lowest vacuum setting that effective-
ly clears the sterile field of surgical smoke.  When
two indicators are illuminated on the suction con-
trol scale, the suction control setting is the value
between the two illuminated percentages.

8. Deactivate the unit by:

a. Pressing the ON/OFF switch on the

membrane panel,

b. Depressing and releasing the footswitch (if

connected), or

c. Releasing either the CUT or COAG button

on the CONMED Smoke Evacuation Pencil
(if Remote Activator is connected).

9. Replace the CONMED 1000SES

TM

Filter when

the Filter Life Remaining Scale flashes RED (0%
Life Remaining). Failure to change the filter will
affect the performance of the system.

2.4 Specifications

Flow Rate:

70 CFM (1982 LPM) with
1-1/4" (32mm) hose
48 CFM (1359 LPM) with
7/8" (22mm) hose

Motor Rating: 115 CFM (3257 LPM)

Static Suction: 97" (246cm) H

2

O

Power:

15 AMPS @ 100V/120V
8 AMPS @ 220/240V

Dimensions:

Height - 14.25" (36.2cm)
Width - 12.38" (34cm)
Length - 18.25" (46.4cm)

Weight:

25.5 lbs (11.6 kg)

Tubing Sizes:

7/8" (22mm), 1-1/4" (32mm),
and 1/4" (6.4mm)

Filter:

CONMED 1000SES

TM

VLSI grade ULPA with carbon

Filtration Efficiency: 99.99995%

Filtration Rate: 0.012 µm

Max. Filter Service Life: 8 Hours

Regulatory Compliance
Units conforming to UL2601-1:
60-6850-001

120 V  60 Hz

60-6850-002

220 V  50 Hz

60-6850-003

240 V  50 Hz

60-6850-004

100 V  50/60 Hz

UL 2601-1 Classifications:
Class 1
Type CF
Ordinary Equipment
Continuous Operation

60-6852-ENG (-).qxd  2/1/99  12:40 PM  Page 2-3

Summary of Contents for 1000 SES

Page 1: ...Service Manual English Version Smoke Evacuation System...

Page 2: ...han intended 1999 CONMED Corporation Mailing Address 310 Broad Street Utica New York 13501 U S A Shipping Address 7211 S Eagle St Englewood Colorado 80112 U S A For Technical Services or Return Author...

Page 3: ...spection 3 1 3 4 Troubleshooting the System 3 2 4 0 CUSTOMER SERVICE 4 1 4 1 Equipment Return 4 1 4 2 Ordering Information 4 1 4 3 Available Accessories 4 1 5 0 THEORY OF OPERATION 5 1 5 1 Filtration...

Page 4: ...1 Control Panel 2 1 2 Back View 2 4 3 Front View 2 4 4 Air Flow Diagram 5 2 5 System Block Diagram 5 3 6 Exploded View 6 5 7 Insulation Assembly 6 6 8 Electrical Schematic 6 7 Figure Title Page...

Page 5: ...subjected to prolonged expo sure The activated carbon used in the CONMED 1000SESTM preferentially removes toxic organic gases rather than water vapor and provides opti mal odor removal The fourth stag...

Page 6: ...CHNICAL PERSONNEL There are no user serviceable components in the CONMED 1000SESTM Smoke Evacuation System Refer service to qualified service per sonnel Use only with the AC cord provided and always p...

Page 7: ...WN arrow button decreas es suction The suction control should be set at the lowest setting to completely remove the surgi cal smoke from the operative site Each time either arrow button is depressed t...

Page 8: ...le signs of structural damage 2 Insert the CONMED 1000SESTM Filter into the round opening in the front of the smoke evac uation system Be sure that the filter is seated completely against the rear of...

Page 9: ...f surgical smoke When two indicators are illuminated on the suction con trol scale the suction control setting is the value between the two illuminated percentages 8 Deactivate the unit by a Pressing...

Page 10: ...Breakers Remote Switch Jack Cautions Product Nameplate I 0 IPX1 Control Panel Filter Inlet Filter CONMED 1000SES Handle Cooling Vents Power Switch Remote Switch Jack Circuit Breakers Cautions Power Co...

Page 11: ...ng it with a cloth that has been dampened not dripping with a disinfectant or a mild soap or detergent solution such as Windex or Formula 409 Windex is a registered trademark of the Drackett Products...

Page 12: ...Tubing 4 Call BioMed or CONMED Tech Services at the numbers listed on the inside front cover of this manual 1 Check power outlet and connection to rear panel of machine 2 Reset circuit breaker on bac...

Page 13: ...001 Remote Switch Adapter 10 Cord Cat No 60 6886 002 Remote Switch Adhesive Pad Bag of 5 Cat No 60 6887 001 CONMED Disposable Smoke Evacuation Handpiece Case of 5 Cat No 60 6800 001 Hand Held Disposa...

Page 14: ...4 2 This page intentionally left blank...

Page 15: ...hose healthcare workers who breathe the gases For this reason the highest quality virgin coal grade activated carbon is used This activated car bon preferentially removes toxic organic gases rather th...

Page 16: ...lled by a triac on the con troller PWB and runs on mains power This motor blower is rated at 115 CFM and a static suction of 97 H2O Replacement brushes for this AC motor are available from CONMED Tech...

Page 17: ...5 3 Figure 5 System Block Diagram...

Page 18: ...5 4 This page intentionally left blank...

Page 19: ...mine the entire system for physical damage Contact BioMed or CONMED for service if damaged Yearly Inspections Perform the chassis ground integrity test per Section 6 3 Check the Service light on the m...

Page 20: ...uum hose clogged 3 Check or replace hose 4 Motor Blower obstructed 4 Contact CONMED Technical Services 2 CONMED 1000 SES 1 Not plugged into electrical 1 Check power outlet or check to does not functio...

Page 21: ...If necessary carefully remove the insulation from the base If any insulation is removed it may not be reinstalled because the adhesive may not func tion properly Newer models do not have adhe sive and...

Page 22: ...luminate after 500 hours of use This is an indication to service the motor brushes Upon inspection if you find the carbon part of the motor brush is long or less replace the brushes Replacement brushe...

Page 23: ...6 5 Figure 6 Exploded View...

Page 24: ...6 6 Figure 7 Insulation Assembly...

Page 25: ...6 7 Figure 8 Electrical Schematic...

Page 26: ...h Jack Wire Harness 1 9 CHP C1X0 8 Microprocessor IC REV 1 1 10 WIRE ASSBLY Wire Assembly w E Terms 1 11 GASKET LED LED Gasket 2 12 MEMBRANEBLNK Membrane Switch Panel Blank 1 13 MEMBRANE CON CONMED Me...