Reading the Filter Life Indicator
Install a CONMED 1000SES
TM
Filter into the
system per the installation instructions contained
in this operator’s manual. When the system is
turned on, the filter life indicator will light up all
the GREEN, AMBER and RED LED’s. This
indicates full filter life. The indicator will
progress to red indicators only as the life of the
filter is exhausted. After eight hours have
elapsed, the RED LED indicator will start blink-
ing to indicate that the filter has expended it use-
ful life and requires replacement.
• CIRCUIT BREAKER (Rear Panel)
Two 15 AMP circuit breakers (8 AMP for inter-
national units) are located on the rear of the unit
They electrically protect both the unit and the
operator from damage or injury. If the unit is
overheated or if there is an electrical surge in the
electrical system, the circuit breakers will “trip”
and the unit will not operate. To restart the sys-
tem, depress the circuit breakers into their origi-
nal position.
When the Service light illuminates, please con-
tact CONMED Technical Services for unit ser-
vice.
2.2 CONMED 1000SES
TM
Filter Instructions
Note: Before installing or removing any filter,
be sure that the system is turned off and the
electrical plug has been disconnected.
Filter Installation Instructions:
The installation of the CONMED 1000SES
TM
Filter into the CONMED 1000SES
TM
Smoke
Evacuation System is quick and simple.
1. Remove the CONMED 1000SES
TM
Filter from
the shipping box and discard any protective wrap-
ping. Examine all filters for damage during ship-
ping and storage. Do not install any filter with
visible signs of structural damage.
2. Insert the CONMED 1000SES
TM
Filter into
the round opening in the front of the smoke evac-
uation system. Be sure that the filter is seated
completely against the rear of the filter chamber.
See Figure 3.
3. Press the FILTER LIFE RESET button on the
membrane control panel for 5 seconds to reset
the filter life clock for the newly installed
CONMED 1000SES
TM
Filter.
WARNING: If fluid is expected to be aspirat-
ed using the CONMED 1000SES
TM
Filter or
the CONMED 1000SES
TM
system, fluid col-
lection devices must be installed with the vac-
uum hose assembly. Failure to install a fluid
collection device may cause filter blockage
and/or electrical damage.
Filter Removal Instructions:
1. After the CONMED 1000SES
TM
Filter has
been exhausted and requires changing, turn the
smoke evacuation system off and disconnect the
unit’s power cord and any accessory tubing
attached to the filter.
2. Pull the CONMED 1000SES
TM
Filter from the
smoke evacuation system and dispose of in accor-
dance with hospital policy. The CONMED
1000SES
TM
Filter may be disposed of or incinerat-
ed.
3. Clean the unit with an appropriate germicide
prior to re-use. Follow the indicated instructions
for maintenance and installation of a new
CONMED 1000SES
TM
Filter.
CAUTION: Using any other filter or accesso-
ry not supplied by CONMED may cause dam-
age to the system and may void the warranty.
WARNING: The CONMED 1000SES
TM
Filter should be changed when the Filter Life
Indicator shows 0% life remaining. Failure to
change this filter may result in decreased effi-
ciency and contamination of the electric
motor, vacuum pump, and sound absorbing
media within the unit.
2.3 Set-up and Operation
The operation of the CONMED 1000SES
TM
Smoke Evacuation System is as follows:
1. Install the CONMED 1000SES
TM
Filter.
60-6852-ENG (-).qxd 2/1/99 12:40 PM Page 2-2
Summary of Contents for 1000 SES
Page 1: ...Service Manual English Version Smoke Evacuation System...
Page 14: ...4 2 This page intentionally left blank...
Page 17: ...5 3 Figure 5 System Block Diagram...
Page 18: ...5 4 This page intentionally left blank...
Page 23: ...6 5 Figure 6 Exploded View...
Page 24: ...6 6 Figure 7 Insulation Assembly...