2.1 System Controls

The electronic system controls on the CONMED
1000SES

TM

are easy to understand and simple to

use. The membrane control panel contains the
power on/off switch, suction power adjustment,
remote activation indicator, filter reset button, fil-
ter life indicator, and service indicator light. See
Figure 1.

Note:  Please be sure to read all instructions
before installing accessories or operating this
equipment.  Failure to do so may result in
damage to the unit and/or personal injury.

• POWER ON/OFF & Standby

There are two ON/OFF switches on the CON-
MED 1000SES

TM

. The ON/OFF switch that

completely disables the electrical power to the
unit is located on the back of the unit.( See
Figure 2.) The ON/OFF switch on the electronic
membrane control panel is located in the upper
right hand corner of the membrane panel. To
power up the machine, place the power switch on
the rear of the machine in the ON position and
press the ON/OFF switch on the membrane
panel until the lights illuminate. Place the unit in
“standby” by pressing the ON/OFF membrane
switch until the  control lights darken. Turn the
unit main power off by placing the power switch
on the back of the unit in the OFF position.

• SUCTION CONTROL (Membrane

Control Panel) See Figure 1.

The amount of suction may be adjusted by
depressing the UP and DOWN Arrow keys.
Depressing the UP arrow button increases suction
and depressing the DOWN arrow button decreas-
es suction. The suction control should be set at
the lowest setting to completely remove the surgi-
cal smoke from the operative site. Each time
either arrow button is depressed, the suction will
increase or decrease by 10%. When two lights are
lit on the flow suction indicator, the actual air
flow is the value between the two illuminated
percentages.

• FOOTSWITCH / REMOTE SWITCH

ACTIVATOR (Membrane Control
Panel) See Figure 1.

A footswitch or a Remote Switch Activator may
be added to the unit by simply plugging a
CONMED activation accessory into the jack on
the back of the unit. When the footswitch is
plugged in, the unit may be turned on or off by
depressing the footswitch pedal once for each
operation. For footswitching operation, the
power switch on the membrane control panel
should be in the ON position. When the
footswitch is plugged in, the light on the mem-
brane control panel entitled REMOTE SWITCH
will light to indicate that the footswitch is active.
For directions on using the Remote Switch
Activator, please see the appropriate section of
this text.

• FILTER LIFE INDICATOR (Membrane
Control Panel) See Figure 1.

The filter life indicator on the membrane control
panel provides a visual indication of the status of
the life of the filter in use. The filter life indicator
for the CONMED 1000SES

TM

is set for eight

hours to correspond to the actual filter life of the
CONMED 1000SES

TM

filter. After each filter

change, depress the RESET button for five sec-
onds to restart the clock for the next filter.

Operating Instructions

Section 2.0

2-1

Figure 1 - Control Panel

60-6852-ENG (-).qxd  2/1/99  12:40 PM  Page 2-1

Summary of Contents for 1000 SES

Page 1: ...Service Manual English Version Smoke Evacuation System...

Page 2: ...han intended 1999 CONMED Corporation Mailing Address 310 Broad Street Utica New York 13501 U S A Shipping Address 7211 S Eagle St Englewood Colorado 80112 U S A For Technical Services or Return Author...

Page 3: ...spection 3 1 3 4 Troubleshooting the System 3 2 4 0 CUSTOMER SERVICE 4 1 4 1 Equipment Return 4 1 4 2 Ordering Information 4 1 4 3 Available Accessories 4 1 5 0 THEORY OF OPERATION 5 1 5 1 Filtration...

Page 4: ...1 Control Panel 2 1 2 Back View 2 4 3 Front View 2 4 4 Air Flow Diagram 5 2 5 System Block Diagram 5 3 6 Exploded View 6 5 7 Insulation Assembly 6 6 8 Electrical Schematic 6 7 Figure Title Page...

Page 5: ...subjected to prolonged expo sure The activated carbon used in the CONMED 1000SESTM preferentially removes toxic organic gases rather than water vapor and provides opti mal odor removal The fourth stag...

Page 6: ...CHNICAL PERSONNEL There are no user serviceable components in the CONMED 1000SESTM Smoke Evacuation System Refer service to qualified service per sonnel Use only with the AC cord provided and always p...

Page 7: ...WN arrow button decreas es suction The suction control should be set at the lowest setting to completely remove the surgi cal smoke from the operative site Each time either arrow button is depressed t...

Page 8: ...le signs of structural damage 2 Insert the CONMED 1000SESTM Filter into the round opening in the front of the smoke evac uation system Be sure that the filter is seated completely against the rear of...

Page 9: ...f surgical smoke When two indicators are illuminated on the suction con trol scale the suction control setting is the value between the two illuminated percentages 8 Deactivate the unit by a Pressing...

Page 10: ...Breakers Remote Switch Jack Cautions Product Nameplate I 0 IPX1 Control Panel Filter Inlet Filter CONMED 1000SES Handle Cooling Vents Power Switch Remote Switch Jack Circuit Breakers Cautions Power Co...

Page 11: ...ng it with a cloth that has been dampened not dripping with a disinfectant or a mild soap or detergent solution such as Windex or Formula 409 Windex is a registered trademark of the Drackett Products...

Page 12: ...Tubing 4 Call BioMed or CONMED Tech Services at the numbers listed on the inside front cover of this manual 1 Check power outlet and connection to rear panel of machine 2 Reset circuit breaker on bac...

Page 13: ...001 Remote Switch Adapter 10 Cord Cat No 60 6886 002 Remote Switch Adhesive Pad Bag of 5 Cat No 60 6887 001 CONMED Disposable Smoke Evacuation Handpiece Case of 5 Cat No 60 6800 001 Hand Held Disposa...

Page 14: ...4 2 This page intentionally left blank...

Page 15: ...hose healthcare workers who breathe the gases For this reason the highest quality virgin coal grade activated carbon is used This activated car bon preferentially removes toxic organic gases rather th...

Page 16: ...lled by a triac on the con troller PWB and runs on mains power This motor blower is rated at 115 CFM and a static suction of 97 H2O Replacement brushes for this AC motor are available from CONMED Tech...

Page 17: ...5 3 Figure 5 System Block Diagram...

Page 18: ...5 4 This page intentionally left blank...

Page 19: ...mine the entire system for physical damage Contact BioMed or CONMED for service if damaged Yearly Inspections Perform the chassis ground integrity test per Section 6 3 Check the Service light on the m...

Page 20: ...uum hose clogged 3 Check or replace hose 4 Motor Blower obstructed 4 Contact CONMED Technical Services 2 CONMED 1000 SES 1 Not plugged into electrical 1 Check power outlet or check to does not functio...

Page 21: ...If necessary carefully remove the insulation from the base If any insulation is removed it may not be reinstalled because the adhesive may not func tion properly Newer models do not have adhe sive and...

Page 22: ...luminate after 500 hours of use This is an indication to service the motor brushes Upon inspection if you find the carbon part of the motor brush is long or less replace the brushes Replacement brushe...

Page 23: ...6 5 Figure 6 Exploded View...

Page 24: ...6 6 Figure 7 Insulation Assembly...

Page 25: ...6 7 Figure 8 Electrical Schematic...

Page 26: ...h Jack Wire Harness 1 9 CHP C1X0 8 Microprocessor IC REV 1 1 10 WIRE ASSBLY Wire Assembly w E Terms 1 11 GASKET LED LED Gasket 2 12 MEMBRANEBLNK Membrane Switch Panel Blank 1 13 MEMBRANE CON CONMED Me...