ConMed 1000 SES Service Manual Download Page 6

CAUTION - Before connecting,
read instructions.

TYPE CF EQUIPMENT CLASS 1

Protect against dripping water.

Alternating Current

1-2

• The CONMED 1000SES

TM

Filters are com-

pletely disposable. Please handle these dispos-
able filters according to your local codes or
regulations and hospital policy. These filters
may be disposed of or incinerated, whichever
is appropriate for your institution.

• The CONMED 1000SES

TM

Filter should be

changed according to the life of the filter. The
CONMED 1000SES

TM

Filter, used with the

CONMED 1000SES

TM

, should not be used

for more than 8 hours. Failure to change the
filter may result in decreased efficiency and
contamination of the electric motor, vacuum
pump, and sound absorbing media within the
unit.

• If fluid is expected to be aspirated by the
CONMED 1000SES

TM

Filter, fluid collection

devices must be installed with the vacuum
hose assembly. Failure to install a fluid collec-
tion device could cause filter blockage and
electrical damage.

WARNING: DO NOT OPERATE THIS
DEVICE IN THE PRESENCE OF FLAM-
MABLE ANESTHETIC GASES,
ENRICHED OXYGEN ATMOSPHERES
OR NITROUS OXIDE.

WARNING: REFER SERVICING TO
QUALIFIED BIOMEDICAL TECHNICAL
PERSONNEL.

There are no user serviceable components in
the CONMED 1000SES

TM

Smoke Evacuation

System. Refer service to qualified service per-
sonnel.

Use only with the AC cord provided and
always plug into a grounded outlet.

The CONMED 1000SES

TM

unit, filters and

accessories are not intended for contact with
patients.

CAUTIONS:

• Do not block either the tubing or the filter.
If either becomes occluded or significantly
restricted, the motor/blower may overheat and
cause the unit to fail.

• Using any other filter or accessory not sup-
plied by CONMED may cause damage to the
system and will void the warranty.

• Care must be exercised in the installation of
hoses, adapters and suction canisters. Failure
to follow the procedures outlined in this
manual may result in overheating of the motor
and may void the unit warranty.

• When using a Suction Canister set up, do
not increase the vacuum setting to more than
50%. Failure to observe this maximum may
result in overheating of the motor and system
failure.

DANGER - Risk of explosion if used
in the presence of flammable
anesthetics.

CAUTION - Grounding reliability can
only be achieved when the equipment is
connected to a receptacle marked
“Hospital Grade” or “Hospital Only”.

CAUTION - Electrical shock hazard.
Do not remove cover. Refer servicing to
qualified service personnel.

IPX1

1.4 Explanation of Symbols

60-6852-ENG (-).qxd  2/1/99  12:40 PM  Page 1-2

Summary of Contents for 1000 SES

Page 1: ...Service Manual English Version Smoke Evacuation System...

Page 2: ...han intended 1999 CONMED Corporation Mailing Address 310 Broad Street Utica New York 13501 U S A Shipping Address 7211 S Eagle St Englewood Colorado 80112 U S A For Technical Services or Return Author...

Page 3: ...spection 3 1 3 4 Troubleshooting the System 3 2 4 0 CUSTOMER SERVICE 4 1 4 1 Equipment Return 4 1 4 2 Ordering Information 4 1 4 3 Available Accessories 4 1 5 0 THEORY OF OPERATION 5 1 5 1 Filtration...

Page 4: ...1 Control Panel 2 1 2 Back View 2 4 3 Front View 2 4 4 Air Flow Diagram 5 2 5 System Block Diagram 5 3 6 Exploded View 6 5 7 Insulation Assembly 6 6 8 Electrical Schematic 6 7 Figure Title Page...

Page 5: ...subjected to prolonged expo sure The activated carbon used in the CONMED 1000SESTM preferentially removes toxic organic gases rather than water vapor and provides opti mal odor removal The fourth stag...

Page 6: ...CHNICAL PERSONNEL There are no user serviceable components in the CONMED 1000SESTM Smoke Evacuation System Refer service to qualified service per sonnel Use only with the AC cord provided and always p...

Page 7: ...WN arrow button decreas es suction The suction control should be set at the lowest setting to completely remove the surgi cal smoke from the operative site Each time either arrow button is depressed t...

Page 8: ...le signs of structural damage 2 Insert the CONMED 1000SESTM Filter into the round opening in the front of the smoke evac uation system Be sure that the filter is seated completely against the rear of...

Page 9: ...f surgical smoke When two indicators are illuminated on the suction con trol scale the suction control setting is the value between the two illuminated percentages 8 Deactivate the unit by a Pressing...

Page 10: ...Breakers Remote Switch Jack Cautions Product Nameplate I 0 IPX1 Control Panel Filter Inlet Filter CONMED 1000SES Handle Cooling Vents Power Switch Remote Switch Jack Circuit Breakers Cautions Power Co...

Page 11: ...ng it with a cloth that has been dampened not dripping with a disinfectant or a mild soap or detergent solution such as Windex or Formula 409 Windex is a registered trademark of the Drackett Products...

Page 12: ...Tubing 4 Call BioMed or CONMED Tech Services at the numbers listed on the inside front cover of this manual 1 Check power outlet and connection to rear panel of machine 2 Reset circuit breaker on bac...

Page 13: ...001 Remote Switch Adapter 10 Cord Cat No 60 6886 002 Remote Switch Adhesive Pad Bag of 5 Cat No 60 6887 001 CONMED Disposable Smoke Evacuation Handpiece Case of 5 Cat No 60 6800 001 Hand Held Disposa...

Page 14: ...4 2 This page intentionally left blank...

Page 15: ...hose healthcare workers who breathe the gases For this reason the highest quality virgin coal grade activated carbon is used This activated car bon preferentially removes toxic organic gases rather th...

Page 16: ...lled by a triac on the con troller PWB and runs on mains power This motor blower is rated at 115 CFM and a static suction of 97 H2O Replacement brushes for this AC motor are available from CONMED Tech...

Page 17: ...5 3 Figure 5 System Block Diagram...

Page 18: ...5 4 This page intentionally left blank...

Page 19: ...mine the entire system for physical damage Contact BioMed or CONMED for service if damaged Yearly Inspections Perform the chassis ground integrity test per Section 6 3 Check the Service light on the m...

Page 20: ...uum hose clogged 3 Check or replace hose 4 Motor Blower obstructed 4 Contact CONMED Technical Services 2 CONMED 1000 SES 1 Not plugged into electrical 1 Check power outlet or check to does not functio...

Page 21: ...If necessary carefully remove the insulation from the base If any insulation is removed it may not be reinstalled because the adhesive may not func tion properly Newer models do not have adhe sive and...

Page 22: ...luminate after 500 hours of use This is an indication to service the motor brushes Upon inspection if you find the carbon part of the motor brush is long or less replace the brushes Replacement brushe...

Page 23: ...6 5 Figure 6 Exploded View...

Page 24: ...6 6 Figure 7 Insulation Assembly...

Page 25: ...6 7 Figure 8 Electrical Schematic...

Page 26: ...h Jack Wire Harness 1 9 CHP C1X0 8 Microprocessor IC REV 1 1 10 WIRE ASSBLY Wire Assembly w E Terms 1 11 GASKET LED LED Gasket 2 12 MEMBRANEBLNK Membrane Switch Panel Blank 1 13 MEMBRANE CON CONMED Me...

Reviews: