26
ENABLING REMOTE SETUP (REPAIR)
The Medical IntelliSwitch™ is equipped with a feature that allows setting of system parameters remotely via the serial port or the
optional Ethernet port. To enable this feature perform the following steps:
1.
Turn AC power OFF to the system.
2.
Open door by loosening the four screws holding it to the cabinet.
3.
Using Figure 15, locate DIP switch SW4-3 at the bottom of the electronic circuit board on the door.
4.
Turn SW4-3 to the ON position.
5.
Close system door and tighten 4 screws holding door to cabinet.
6.
Turn AC power back ON. The Auxiliary Gas Supply Option will be enabled on power ON.
Note:
When this feature is enabled, system setup parameters are saved to internal memory in the control electronics. The DIP switches on
the control board for these setup parameters will be ignored. To allow use of the setup parameter DIP switches on the control board,
SW4-3 must be in the OFF position.
Figure 15