22
RESETTING UNITS OF MEASURE (REPAIR)
The Medical IntelliSwitch™ can display pressure in three different units of measure, PSI, BAR, and MPA. The factory default is PSI.
To change units of measure perform the following steps:
1.
Turn AC power OFF to the system.
2.
Open door by loosening the four screws holding it to the cabinet.
3.
Using Figure 11, locate DIP switch SW2 at the bottom of the electronic circuit board on the door.
4.
Using Table 6, set the appropriate units of measure.
5.
Close system door and tighten 4 screws holding door to cabinet.
6.
Turn AC power back ON. The new units of measure will take effect on power ON.
Figure 11
7.
The indicator light son the front panel show the units of measure selected. If the PSI and BAR indicator lights are OFF, the
system is displaying MPA units.
Table 6
Comm
Mode
SW2-1 SW2-2
PSI OFF
OFF
default
BAR
ON
OFF
MPA OFF
ON