If the expected results are obtained the HT-PR is functioning properly and that it is safe to use. If
the results of the performance test are not in accordance with the above, service is required.
Remove the HT-PR from service and contact Compliance West USA, Inc. for servicing
information.
Figure 6. Expected results for performance test
Insulation
Leakage current
Ground Switch
R value
Test Result
Good
I < I Setting
OFF
X
A
X
X
ON
R >= R Setting
B
Good
I < I Setting
ON
R <= R Setting
C
Good
I >= I Setting
OFF
R <= R Setting
D
Weak
I < I Setting
OFF
R <= R Setting
E
Table 5. Test Conditions and results
Figure 7. Breakdown Fail test (Not recommended).