27
The small size of the
Ground Check
switch will allow a calibration-type adhesive label to be
placed over it to allow for tamper-proof operation.
Operation Techniques
The following paragraphs describe how to operate your HT-PR Dielectric Withstand Tester.
Before proceeding with testing, the HT-PR tester should be set up for production line testing.
See Section above before continuing.
CAUTION:
High voltage is generated by the HT-PR. Although the chassis of the equipment under test
is grounded by the HT-PR, a risk of shock exists. Exercise care when using the HT-PR.
Daily Operation Test
The operation of the HT-PR should be audited daily by conducting the tests described in the
Initial Checkout Procedure
section of this Manual. Compliance West USA also offers the
models HTT-1 and HTT-1R, which can be used to quickly verify the operation of your HT-PR.
Contact the factory for details.
Connections
This section describes the most common configuration for testing different devices (3 wire and 2
wire testing), the description of the output receptacle and what are the different measure points
(Ground, leakage and breakdown).
HT-PR output receptacle Description.
The following figure shows the output receptacle of the HT-PR and the internal connections, the
high voltages goes thru the
line/neutral
orange receptacle to the
return
connector of the front
panel of the tester, in the HT-PR a red NEMA 5-15R receptacle are include that is connected to
the High voltage (line/neutral of the orange receptacle) for different test cables, the ground
resistance is measured from the ground pin of the orange receptacle to the return connector.
Figure 13. Connection and description of output receptacle