11
Specifications
ELECTRICAL
Output
0 - 2000 Volts AC HT-2000PR, HT-3000PR
0 - 2800 Volts DC HT-2800PR, HT-3000PR
Pigtail and Receptacle adapters available for most
configurations. Contact Compliance West for more
information.
Leakage Current
0-20mA AC HT-2000PR, HT-3000PR
0-5mA DC HT-2800PR, HT-3000PR
Pass/Fail Criteria:
Continuity:
Pass/Fail Point user adjustable 0.01-1.50
Ω
±
0.02
Ω
(Defeatable).
Leakage Current:
Pass/Fail point user adjustable.
Dielectric Breakdown:
Separate high frequency detection circuit for
breakdown spike detection
Test Time:
User adjustable 1-60 sec.
Voltage Ramp-up Time:
User adjustable 1-5 sec.
Pass/Fail Repeatability
±
3%
Meter accuracy
+ 2% from 500V to maximum limit voltage
Duty cycle
100 %
Test adjustments
Rear Panel:
Ramp Time
Test Time
Leakage Limit
Voltage Adjust
Ground Offset
Tamper proof operation available by removing
adjustment shafts.
ENVIRONMENTAL
Operating Temperature
15-40
°
C
Relative Humidity Range
0-90% non-condensing
GENERAL
Input power requirements
See rear panel for input voltage and fuse ratings
Weight
10½ lbs.
SAFETY AGENCY TOPICS
Transformer Output
< 500VA
Visual Indication of Voltage Output
Provided by front panel meter, directly connected to
high voltage output
Failure Indication
Audible, provided by internal buzzer
Visual, provided by red LEDs on front panel
Test automatically terminated on failure
Continuity Test
Provided; pass/fail point adjustable. May be defeated
for testing of products with two wire power supply
cords.
Leakage Test
Provided; 5mA AC/2mA DC factory set pass/fail
point, user adjustable.
Operation Instructions
Provided on rear panel.
Table 1. HT-PR Specifications