31
Section 5
Maintenance and Calibration
WARNING
THESE SERVICE INSTRUCTIONS ARE FOR USE BY QUALIFIED PERSONNEL
ONLY. TO AVOID ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING
OTHER THAN THAT CONTAINED IN THE OPERATING INSTRUCTIONS UNLESS
YOU ARE QUALIFIED TO DO SO.
Introduction
This section of the manual contains maintenance information for the HT-PR Dielectric
Withstand Tester. This maintenance information is divided into service information, general
maintenance, a performance test, and a calibration procedure. The performance test is
recommended as an acceptance test when the instrument is first received, and later as a
preventative maintenance tool to verify proper instrument operation. A 1-year calibration cycle
is recommended to maintain the specifications given in Section 1. A three inch piece of approx.
18 AWG insulated hookup wire with each end stripped back ½" and a lead with a banana plug on
one end are required to conduct the performance test. The test equipment required for the
calibration procedure is a DMM able to read true rms 0-2000Vac
±
1%, 0-2800Vdc
±
1%, 0.2
Ω
and 1.5
Ω
1/4 w 1% resistor.
Service Information
The HT-PR is warranted to the original purchaser for a period of 1 year. This warranty does not
cover problems due to misuse or neglect.
Malfunctions which occur within the limits of the warranty will be corrected at no charge. Mail
the instrument post paid to the manufacturer. Dated proof of purchase is required for all in-
warranty repairs.
The manufacturer is also available for calibration and/or repair of instruments that are beyond
their warranty period. Contact the manufacturer for a cost quotation. Ship the instrument and
your remittance according to the instructions given by the manufacturer.