6.  Push  and  hold  the  Test  Button.    (Holding  the  test  button  will  continue  the  test  until  the  button  is 

released.) 

7.  With  the  Test  Button  pressed,  turn  the  screwdriver until both the red and green lights are on.  The 

internal buzzer may sound at the calibration point. 

8.  Now, set the duration of the test.  As shipped from the factory, the GF-30AC-T will only test while 

the test button is pressed.  By inserting the plastic calibration screwdriver into the potentiometer into 
the  potentiometer  marked  "DURATION",  the  test  time  can  be  set.    A  time  of  approx.  5  seconds  is 
recommended, subject to approval by safety agency inspection personnel.

 

 

Time Duration Adjustment 

 

1. 

Connect your GF-30AC-T Tester to a correctly rated source of supply and turn the power switch to 
the I or ON position.  The tester will perform a self-diagnostic test.  

2. 

Insert  the  plastic  calibration  screwdriver  into  the  potentiometer  behind  the  lower  hole  in  the  rear 
panel marked "DURATION". 

3. 

Push  and  release  the  Test  Button.    The  test  duration  is  counted  from  the  RELEASE  of  the  Test 
Button and is controlled by the setting of the DURATION control.  Please note that if a manually 
timed test is conducted, the test will continue for the amount of time set during this Procedure after 
the Test Button is released.  Therefore, we recommend that the Test Duration control be adjusted to 
its minimum position when a manual test is conducted. 

4. 

Adjust the control with the calibration screwdriver until the desired test duration is set.

 

 

Operating Techniques 

 

The  following  paragraphs  describe  how  to  operate  your  GF-30AC-T  Tester.  In the following sections, 
EUT means Equipment Under Test. 
 

Setting the Test Time Duration 

 

1. 

Set  the  test  duration  as  shown  in  "Time  Duration  Adjustment"  to  ensure  the  production  line  test 
will be done for the desired amount of time.  A test time of 5 seconds is recommended, subject to 
acceptance by safety agency inspection personnel. 

 

Testing the EUT 

 
This section describes how the Tester is used to conduct an actual test. 
  

CAUTION 

High current.  Risk of burns.  Remove any conductive jewelry before using the tester. 

 

NOTE 

If the EUT uses a detachable power cord, use the Test Power Cord packed with the instrument, not 

the EUT power cord.  The accuracy of the test procedure depends on use of the same power cord 

type for every test.

 

 

1.  Connect your Tester to a correctly rated source of supply and turn the power switch to the I or ON 

position.    The  tester  will  perform  a  self-diagnostic  test.    Make  sure  the  buzzer  sounds  and  the  red 
light is illuminated during this test.  

Summary of Contents for GF-30AC-T

Page 1: ...GF 30AC T Ground Circuit Tester Instruction Manual...

Page 2: ...represents the latest in high current production line testing To fully appreciate all the features of your new meter we suggest that you take a few moments to review this manual Compliance West USA s...

Page 3: ...3...

Page 4: ...Duration Adjustment 12 Operating Techniques 12 Setting the Test Time Duration 12 Testing the EUT 12 Test results 13 Green Light 13 Red Light 13 Red Light and Buzzer Sound 13 Simultaneous Red and Gree...

Page 5: ...etermined amount of time after the Test Button has been released The instrument advises the operator of any internal malfunction affecting the output A one time setup procedure allows the instrument t...

Page 6: ...est Duration Timer 0 25 sec approx Duty cycle 100 ENVIRONMENTAL Operating Temperature 15 35 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements See rear panel for input volt...

Page 7: ...rotection is provided to prevent damage during shipment We recommend that the instrument be surrounded by at least one inch of shock absorbing material on all sides of the container Remove the Tester...

Page 8: ...Red Fail Indicator Lamp Incandescent lamp When lit indicates that equipment under test has more than 100 milliohms in grounding path or is incorrectly connected to tester Replace with 6 3V type 86 la...

Page 9: ...section of this manual for procedure 5 Appliance Inlet Fuseholder Power Switch Use 18 AWG cordset to connect tester to appropriate source of supply Replace line fuse Turn tester ON OFF 6 Fuse replacem...

Page 10: ...ductive surface Push the Test button The GF 30AC T should indicate a Fail by lighting the red light and sounding the internal buzzer If your Tester yielded the proper results it is working properly If...

Page 11: ...safety agency inspection personnel Adjusting the Pass Fail Point for products with non removable power cords Use this adjustment procedure if the equipment under test uses a non detachable power cord...

Page 12: ...t will continue for the amount of time set during this Procedure after the Test Button is released Therefore we recommend that the Test Duration control be adjusted to its minimum position when a manu...

Page 13: ...the ground system of the EUT has less than 100 milliohm resistance For most categories of equipment this result is acceptable Red Light A red light indicates the ground system of the EUT has more than...

Page 14: ...er operation by following steps 1 3 in Adjusting the Pass Fail Point using the supplied Test Cordset Press the Test Button and measure the voltage across the calibration resistor If the unit is workin...

Page 15: ...st USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego CA 921...

Page 16: ...The Test equipment required for both the performance test and calibration is a DMM able to read true rms 0 3 VAC 1 Service Information The GF 30AC T is warranted to the original purchaser for a perio...

Page 17: ...iagram for high voltage areas Cleaning CAUTION Do not use aromatic hydrocarbons or chlorinated solvents for cleaning These solutions will react with the plastic materials used in the instrument Clean...

Page 18: ...el calibration resistor using the alligator clip wire assembly 3 Connect the DMM across the calibration resistor 4 Push and hold the Test Button on the front panel For a 25 amp output the DMM should r...

Page 19: ...libration resistor is 2 5 Vrms Using Ohms law this corresponds to an output current of 25 Amps Other output current levels can be set in the same manner Adjust Buzzer 5 With the calibration resistor s...

Reviews: