Adjusting the Pass/Fail Point using the supplied Test Cordset 

 

Use this adjustment procedure if the equipment under test uses a detachable power cord.  Always use the 
test cord supplied with the GF-30AC-T.  Do not use the cordset of the EUT for this test. 
1. 

Connect your GF-30AC-T Tester to a correctly rated source of supply and turn the power switch to 
the I or ON position. 

2. 

Plug  the  split  end  plug  of  the  10  AWG  Test  return  lead  into  the  Return  Lead  Receptacle  of  the 
Tester.  Connect the other end to the calibration resistor bolted to the rear panel. 

3. 

Plug  the  Test  Power  Cord  into  the  Test  Power  Receptacle  of  the  Tester.    Attach  the  alligator 
clip/wire  assembly  to  the  center  lead  at  the  end  of  the  Test  Power  Cord  and  to  the  calibration 
resistor. 

4. 

Adjust duration to minimum. See “Time duration adjustmen” on the next page. 

5. 

Insert  the  plastic  calibration  screwdriver  into  the  potentiometer  on  the  rear  panel  marked  "CAL 
POINT".   

6. 

Push  and  hold  the  Test  Button.    (Holding  the  test  button  will  continue  the  test until the button is 
released.) 

7. 

With the Test Button pressed, turn the screwdriver until both the red and green lights are on.  The 
internal buzzer may sound at the calibration point. 

8. 

Now, set the duration of the test.  As shipped from the factory, the GF-30AC-T will only test while 
the  test  button  is  pressed.    By  inserting  the  plastic  calibration  screwdriver  into  the  potentiometer 
marked  "DURATION",  the  test  time  can  be  set.    A  time  of  approximately  5  seconds  is 
recommended, subject to approval by safety agency inspection personnel.

 

 

Adjusting  the Pass/Fail Point for products with non-removable power cords 

 

Use  this  adjustment  procedure  if  the  equipment  under  test  uses  a  non-detachable  power  cord.  (A non-
detachable power cord is permanently connected to the EUT with a strain relief bushing.) 
 

NOTE 

Ensure that the non-detachable power cord meets the length and conductor size requirements of Table 1-

2 before attempting to calibrate the tester.  If a non-detachable power cord not meeting Table 1-2 is used, 

the tester calibration procedure may not be successful.  However, if the calibration procedure is 

successful, the GF-30AC-T will test products correctly. 

 
1.  This procedure requires the use of a power cord assembly from the EUT.  Obtain one from stock and 

trim it to the total length used in the EUT, making sure to include the length of the cord inside the 
enclosure  of  the  EUT.    Crimp  or  solder  an  alligator  clip  to  the  green  or  green  with  yellow  stripes 
(grounding) conductor of the cord at the open end. 

2.  Connect your GF-30AC-T Tester to a correctly rated source of supply and turn the power switch to 

the I or ON position.  The tester will perform a self-diagnostic test.  Make sure the buzzer sounds and 
the red light is illuminated during this test. 

3.  Plug  the  split  end  plug  of  the  10  AWG  Test  return  lead  into  the  Return  Lead  Receptacle  of  the 

Tester.  Connect the other end to the calibration resistor bolted to the rear panel. 

4.  Plug  the  EUT  Power  Cord  prepared  in  Step  1  above  into  the  Test  Power  Receptacle  of  the  Tester.  

Attach its alligator clip to the other lead of the calibration resistor. 

5.  Insert  the  plastic  calibration  screwdriver  into  the  potentiometer  on  the  rear  panel  marked  "CAL 

POINT".   

Summary of Contents for GF-30AC-T

Page 1: ...GF 30AC T Ground Circuit Tester Instruction Manual...

Page 2: ...represents the latest in high current production line testing To fully appreciate all the features of your new meter we suggest that you take a few moments to review this manual Compliance West USA s...

Page 3: ...3...

Page 4: ...Duration Adjustment 12 Operating Techniques 12 Setting the Test Time Duration 12 Testing the EUT 12 Test results 13 Green Light 13 Red Light 13 Red Light and Buzzer Sound 13 Simultaneous Red and Gree...

Page 5: ...etermined amount of time after the Test Button has been released The instrument advises the operator of any internal malfunction affecting the output A one time setup procedure allows the instrument t...

Page 6: ...est Duration Timer 0 25 sec approx Duty cycle 100 ENVIRONMENTAL Operating Temperature 15 35 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements See rear panel for input volt...

Page 7: ...rotection is provided to prevent damage during shipment We recommend that the instrument be surrounded by at least one inch of shock absorbing material on all sides of the container Remove the Tester...

Page 8: ...Red Fail Indicator Lamp Incandescent lamp When lit indicates that equipment under test has more than 100 milliohms in grounding path or is incorrectly connected to tester Replace with 6 3V type 86 la...

Page 9: ...section of this manual for procedure 5 Appliance Inlet Fuseholder Power Switch Use 18 AWG cordset to connect tester to appropriate source of supply Replace line fuse Turn tester ON OFF 6 Fuse replacem...

Page 10: ...ductive surface Push the Test button The GF 30AC T should indicate a Fail by lighting the red light and sounding the internal buzzer If your Tester yielded the proper results it is working properly If...

Page 11: ...safety agency inspection personnel Adjusting the Pass Fail Point for products with non removable power cords Use this adjustment procedure if the equipment under test uses a non detachable power cord...

Page 12: ...t will continue for the amount of time set during this Procedure after the Test Button is released Therefore we recommend that the Test Duration control be adjusted to its minimum position when a manu...

Page 13: ...the ground system of the EUT has less than 100 milliohm resistance For most categories of equipment this result is acceptable Red Light A red light indicates the ground system of the EUT has more than...

Page 14: ...er operation by following steps 1 3 in Adjusting the Pass Fail Point using the supplied Test Cordset Press the Test Button and measure the voltage across the calibration resistor If the unit is workin...

Page 15: ...st USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego CA 921...

Page 16: ...The Test equipment required for both the performance test and calibration is a DMM able to read true rms 0 3 VAC 1 Service Information The GF 30AC T is warranted to the original purchaser for a perio...

Page 17: ...iagram for high voltage areas Cleaning CAUTION Do not use aromatic hydrocarbons or chlorinated solvents for cleaning These solutions will react with the plastic materials used in the instrument Clean...

Page 18: ...el calibration resistor using the alligator clip wire assembly 3 Connect the DMM across the calibration resistor 4 Push and hold the Test Button on the front panel For a 25 amp output the DMM should r...

Page 19: ...libration resistor is 2 5 Vrms Using Ohms law this corresponds to an output current of 25 Amps Other output current levels can be set in the same manner Adjust Buzzer 5 With the calibration resistor s...