9
Figure 2-2. Control, Indicators, Connectors - Model GF-30AC-T Rear Panel
ITEM NO.
NAME
FUNCTION
1
Legend
Truth table indication of results.
2
Directions
Provides directions for tester operation to line test
personnel.
3
CAL Adjustment
Used to adjust PASS/FAIL point. See "Calibrating
the GF-30AC-T Tester" section of this manual for
procedure.
4
Duration Adjustment
Used to adjust automatic test duration. See
"Calibrating the GF-30AC-T Tester" section of this
manual for procedure.
5
Appliance Inlet / Fuseholder
/ Power Switch
Use 18 AWG cordset to connect tester to
appropriate source of supply. Replace line fuse.
Turn tester ON/OFF.
6
Fuse replacement warning /
Rating of supply
Specifies replacement fuse and supply voltage used.
7
Calibration Resistor
0.1 50W for resistor used in the Adjustment of
the Pass/Fail point.
8
Test Link Port
9-pin D subminiature female connector for remote
control with a dielectic withstand tester (Only
Compliance West Models).
9
Thermal Cooling Fan
Required to avoid component overheat.
Table 2-2. Control, Indicators, Connectors - Model GF-30AC-T Rear Panel