Initial Checkout Procedure 

 

The  following  procedure  will  allow  you  to  verify  that  the  Tester  is  working  correctly  before  use.    The 
only test equipment required is the unit itself, the 10 AWG Test Return Lead, the 14 AWG Test Cord, 
and the alligator clip/wire assembly. 

 

CAUTION 

High  current.    Risk  of  burns.    Remove  any  conductive  jewelry  before  using  the 
Tester. 

 

1. 

Connect your Tester to a correctly rated source of supply, remove any other test leads, and turn 
the  power  switch  to  the  I  or  ON  position.    The  tester  will  perform  a self-diagnostic test.  Wait 
until the red and/or green lights are both out before proceeding. Confirm that the blue test button 
is illuminated.  If it is not, check the source of supply. 

2. 

Plug the split end plug of the 10 AWG Test return lead into the Return Lead Receptacle of the 
Tester.    

3. 

Plug  the  Test  Power  Cord  into  the  Test  Power  Receptacle  of  the  Tester.    Attach  the  alligator 
clip/wire assembly to the center lead of the IEC connector on the exposed end of the Test Power 
Cord.  Connect the two Test cords together. 

4. 

Push the Test Button.  The GF-30AC-T should indicate a "Pass" by lighting the green light. 

5. 

Disconnect  the Test leads from each other and carefully lay them on a non-conductive surface.  
Push  the  Test  button.    The  GF-30AC-T  should  indicate  a  "Fail"  by  lighting  the  red  light  and 
sounding the internal buzzer. 

 
If your Tester yielded the proper results, it is working properly.  If it did not, call Compliance West USA 
for assistance. 

Adjusting the Pass/Fail Point 

 

This  adjustment  allows  the  GF-30AC-T  to  compensate  for  the  resistance  of  the  cordset  and  test  return 
lead.  If done properly, it allows the GF-30AC-T to measure only the resistance of the ground circuit of 
the EUT, while compensating for cord resistance.   
It is necessary to check the calibration point each time a different length or gauge Test Lead or Cord is 
used, and every time the instrument is returned to service following an annual calibration.  The 25 amp 
output level is not adjustable.  In the following Section, EUT means Equipment Under Test. 

 

CAUTION 

High current.  Risk of burns.  Remove any conductive jewelry before using the Tester. 

 

NOTE 

The duty cycle of the rear-panel mounted calibration resistor is 20%, maximum 5 seconds on, let it 

cool down for 15 seconds, to improve the duty cycle, mounting it to a big heatsink and using a fan 

can at least double the dutty cycle ratings.  Failure to adhere to this duty cycle may damage the 

calibration resistor. 

 

NOTE 

This adjustment procedure sets the tester Pass/Fail indication point only.  It does not take the place 

of the required annual calibration required for the GF-30AC-T.

 

 
 

Summary of Contents for GF-30AC-T

Page 1: ...GF 30AC T Ground Circuit Tester Instruction Manual...

Page 2: ...represents the latest in high current production line testing To fully appreciate all the features of your new meter we suggest that you take a few moments to review this manual Compliance West USA s...

Page 3: ...3...

Page 4: ...Duration Adjustment 12 Operating Techniques 12 Setting the Test Time Duration 12 Testing the EUT 12 Test results 13 Green Light 13 Red Light 13 Red Light and Buzzer Sound 13 Simultaneous Red and Gree...

Page 5: ...etermined amount of time after the Test Button has been released The instrument advises the operator of any internal malfunction affecting the output A one time setup procedure allows the instrument t...

Page 6: ...est Duration Timer 0 25 sec approx Duty cycle 100 ENVIRONMENTAL Operating Temperature 15 35 C Relative Humidity Range 0 90 non condensing GENERAL Input power requirements See rear panel for input volt...

Page 7: ...rotection is provided to prevent damage during shipment We recommend that the instrument be surrounded by at least one inch of shock absorbing material on all sides of the container Remove the Tester...

Page 8: ...Red Fail Indicator Lamp Incandescent lamp When lit indicates that equipment under test has more than 100 milliohms in grounding path or is incorrectly connected to tester Replace with 6 3V type 86 la...

Page 9: ...section of this manual for procedure 5 Appliance Inlet Fuseholder Power Switch Use 18 AWG cordset to connect tester to appropriate source of supply Replace line fuse Turn tester ON OFF 6 Fuse replacem...

Page 10: ...ductive surface Push the Test button The GF 30AC T should indicate a Fail by lighting the red light and sounding the internal buzzer If your Tester yielded the proper results it is working properly If...

Page 11: ...safety agency inspection personnel Adjusting the Pass Fail Point for products with non removable power cords Use this adjustment procedure if the equipment under test uses a non detachable power cord...

Page 12: ...t will continue for the amount of time set during this Procedure after the Test Button is released Therefore we recommend that the Test Duration control be adjusted to its minimum position when a manu...

Page 13: ...the ground system of the EUT has less than 100 milliohm resistance For most categories of equipment this result is acceptable Red Light A red light indicates the ground system of the EUT has more than...

Page 14: ...er operation by following steps 1 3 in Adjusting the Pass Fail Point using the supplied Test Cordset Press the Test Button and measure the voltage across the calibration resistor If the unit is workin...

Page 15: ...st USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego CA 921...

Page 16: ...The Test equipment required for both the performance test and calibration is a DMM able to read true rms 0 3 VAC 1 Service Information The GF 30AC T is warranted to the original purchaser for a perio...

Page 17: ...iagram for high voltage areas Cleaning CAUTION Do not use aromatic hydrocarbons or chlorinated solvents for cleaning These solutions will react with the plastic materials used in the instrument Clean...

Page 18: ...el calibration resistor using the alligator clip wire assembly 3 Connect the DMM across the calibration resistor 4 Push and hold the Test Button on the front panel For a 25 amp output the DMM should r...

Page 19: ...libration resistor is 2 5 Vrms Using Ohms law this corresponds to an output current of 25 Amps Other output current levels can be set in the same manner Adjust Buzzer 5 With the calibration resistor s...

Reviews: