INSTRUCTION MANUAL
SGTEL-168
TELECOM LINE SURGE GENERATOR
1912 1 E l T o ro R d
●
Si l vera do, Ca lifo rni a 9 26 76
●
(9 49) 459 -96 00
●
co m-p o wer. com
REV070516
4.4.1.1 Type A – 800V Surge (metallic) Terminal Connections
This surge waveform is intended to be applied between any
pair of connections on which lightning surges may occur,
including (for example):
a)
tip to ring;
b)
tip-1 to ring-1; and,
c)
for a 4-wire connection that uses simplex signaling, tip to
ring-1 and ring to tip-1.
The SGTEL-168 applies this surge between Terminals (A) and (B).
Therefore, the EUT lines to be tested should be connected to
these terminals for this surge, as shown in Figure 8.
FIGURE 8 -
Coupling Example for
Type A – 800V (metallic) Surge
chassis
chassis
phone
phone
line
line
EQUIPMENT
EQUIPMENT
UNDER
UNDER
TEST
TEST
(EUT)
(EUT)
TIP
RING
GND