4.4.1.4
Type B – 1500V Surge (longitudinal) Terminal Connections
This surge waveform is intended to be applied to any pair of
connections on which lightning surges may occur. This includes
(for example):
d)
tip/ring pair to earth;
e)
tip-1/ring-1 pair to earth; and,
f)
to all leads intended for connection to non-approved
equipment, connected together.
The SGTEL-168 applies this surge from Terminals (A) and (B) to
the Ground Terminal (G). Therefore, the EUT lines to be tested
should be connected to these terminals for this surge, as shown
in Figure 11.
FIGURE 11 -
Coupling Example for
Type B – 1500V (longitudinal) Surge
chassis
chassis
phone
phone
line
line
EQUIPMENT
EQUIPMENT
UNDER
UNDER
TEST
TEST
(EUT)
(EUT)
TIP
RING
GND