
CI-154 User’s Manual
Page 36
2.13.9
Printing a GMP 2008 report
The GMP 2008 report screen is accessed from the report menu screen. Next select
[GMP 2008] to enable GMP 2008 reports. Press [
SET CLASS
]; from the screen that
follows, press the desired GMP 2008 grade for the area being monitored. The
selected key will turn green when pressed. The at size keys are always selected
since GMP 2008 reports always include both sizes. Press [
EXIT
] to return to the
report menu screen.
Next, press the [
ISO
/209
E
/
GMP
] key, and then press the [
] key at the bottom of
the following screen to print the GMP 2008 report. Pass/fail criteria for a GMP 2008
report are given in the table below.
Note that minimum sample volume requirements of GMP 2008 are not considered in
the Pass/Fail criteria. The user is responsible for determining that sampling volume,
proper sampling locations and other requirements such as probe positions are met.
The minimum number of sampling locations may be enforced by entering the area of
the cleanroom when setting the desired class. The report will always include the
program name, and will include the Unit ID if one exists. The report will also include
user ID’s if security is enabled.
Grade
Maximum permitted number of particles/m
3
equal to or above
At rest
In operation
0.5 µm
5.0 µm
0.5 µm
5.0 µm
A
3520
20
3520
20
B
3520
29
352,000
2,900
C
352,000
2,900
3,520,000
29,000
D
3,520,000
29,000
Not defined
Not defined
Summary of Contents for CI-154
Page 2: ...CI 154 User s Manual Page ii ...
Page 4: ...CI 154 User s Manual Page iv ...
Page 9: ...CI 154 User s Manual Page ix FRONT VIEW OF CI 154 ...
Page 10: ...CI 154 User s Manual Page x REAR PANEL OF CI 154 ...
Page 16: ...CI 154 User s Manual Page 6 ...
Page 18: ...CI 154 User s Manual Page 8 ...
Page 80: ...CI 154 User s Manual Page 70 ...
Page 84: ...CI 154 User s Manual Page 74 ...
Page 92: ...CI 154 User s Manual Page 82 ...
Page 96: ...CI 154 User s Manual Page 86 ...
Page 98: ...CI 154 User s Manual Page 88 ...