CI-154 User’s Manual
Page 7
1.5
Specifications
Illumination source: Laser diode, 50 mW
Light Collection Optics: Elliptical mirror
Detection: Silicon photo diode
Size sensitivity: 0.30
m
Particle channel sizes: Four sizes: 0.3, 0.5, 1.0, 5.0
m
Size resolution: Meets requirements of ISO 21501-4 - 10% typical
Concentration limit: 1.0 x 10
6
particles per cubic foot (10% coincidence)
(35.3 x 10
6
particles per cubic meter)
Sample flow rate: 1.00 cubic feet per minute
(0.028 cubic meter or 28.3 liters per minute)
Flow rate control: Electronic, automatic closed loop
Sample tube extension: 3/8” (9.525 mm) I.D., up to 20 ft. (6.1 meters) max. length
Max. False Count Rate: 7.1 per CM >0.3 µm (95% UCL) per ISO 21501-4
(1 count per five minutes)
Battery: Replaceable 9.6 Volts, 12.6 Amp hour Nickel Metal Hydride
Battery operation time: 5.5 to 6.5 hours continuous operation
Time to charge battery:3.5 to 4.0 hours
Display: TFT Color LCD, ¼ VGA touch screen
Printer: Thermal Dot Matrix
Memory: 3,000 samples, date and time stamped
Unit ID: User selectable
User ID: Up to 20 users, besides supervisor, with password
Security: Four security levels with user authentication
Location ID: Up to 100 per program; 16 characters long.
Program: Up to 30 user-defined and named, containing IDs and settings.
Input/Output (RS-232): ASCII command set, CSV data output with check code
+5 VDC @ 400 mA provided on pin 9
Input/Output (Ethernet) 10/100Base-T, half or full duplex (auto sensing), http web
server, Ethernet Modbus
®
server and TELNET server
Input/Output (USB) Transfer of stored data memory to jump drive
Transfer of user-defined programs to/from jump drive
Transfer of unit settings to/from unit via jump drive
Beep On Count: 0.3
m and greater.
Audible alarm: Built-in
Alarm output: Dry contacts, normally open or closed via alarm connector.
Connector also has +12VDC @ 250mA.
Alarm limits: Programmable for all particle channels and flow.
Dimensions: H-W-D: 8.65 x 8.5 x 9.5 inches (21.97 X 21.6 X 24.1 cm)
Weight: 11.0 lb (4.99 Kg) without battery
14.25 lb (6.46 Kg) with battery
Power: Auto switching: 100–240 VAC, 47/63 Hz, 90VA max
Fuses: two 5 X 20 mm, 250V/2.0 Amp Slow
Operating range: 0–36º C, 10–90% RH, non-condensing
Calibration: Factory Prime Calibration complying with ISO 21501-4.
NIST traceable using CalPro 2.00 (or later) software program.
Calibration frequency: Recommended every 12 months.
Gas compatibility: Air, nitrogen; consult factory regarding other gases.
Summary of Contents for CI-154
Page 2: ...CI 154 User s Manual Page ii ...
Page 4: ...CI 154 User s Manual Page iv ...
Page 9: ...CI 154 User s Manual Page ix FRONT VIEW OF CI 154 ...
Page 10: ...CI 154 User s Manual Page x REAR PANEL OF CI 154 ...
Page 16: ...CI 154 User s Manual Page 6 ...
Page 18: ...CI 154 User s Manual Page 8 ...
Page 80: ...CI 154 User s Manual Page 70 ...
Page 84: ...CI 154 User s Manual Page 74 ...
Page 92: ...CI 154 User s Manual Page 82 ...
Page 96: ...CI 154 User s Manual Page 86 ...
Page 98: ...CI 154 User s Manual Page 88 ...