91
IEC 61010
IEC 61557
MADE IN FRANCE
10.4. MEtRoloGICAl CHECk
like all measuring or testing devices, the instrument must be checked regularly.
This instrument should be checked at least once a year. For checks and calibrations, contact one of our accredited metrology labo-
ratories (information and contact details available on request), at our Chauvin Arnoux subsidiary or the branch in your country.
10.5. REPAIR
For all repairs before or after expiry of warranty, please return the device to your distributor.
10.6. UPDAtInG of tHE IntERnAl softwARE
With a view to providing, at all times, the best possible service in terms of performance and technical upgrades, Chauvin Arnoux
invites you to update the embedded software of the device by downloading the new version, available free of charge on our web site.
Our site:
http://www.chauvin-arnoux.com
Sign in and open your account.
Then go to “Software support space”, then “Freely available software”, then “C.A 6116”.
Set the switch to Z
S
(R
A
/S
EL.
).
SET UP
OFF
Press the 3 keys indicated below simultaneously.
SET UP
OFF
OK
7. Put the battery compartment cover back in place and screw the two quarter-turn screws back.
8. Charge the new battery
fully
before using the device.
9. If the battery remained disconnected for more than 5 minutes, you may have to reprogram the date and time of the device
(see §5).
Attention :
Whenever the battery is disconnected,
even if it is not replaced
, it must be charged fully. This is so that the device
will know the charge condition of the battery (this information is lost when it is disconnected).
10.3. REsEttInG tHE DEvICE
If the device crashes, it can be reset, like a PC.
Summary of Contents for C.A 6116
Page 1: ...C A6116 Installation TESTER E N G L I S H User s manual...
Page 95: ...95...