To see the next display page.
INSULATION
02
/
11
/
2009 10
:
47
500 k
Ω
- - .- Hz
U
M
Ω
0 . 0 V
../..
2 %
Press TEST until the measurement
is stable
External voltage present on the
terminals just before the start of the
measurement.
Use the key to return to the previ-
ous display page.
For help with connections or any other information, use the on-line help.
3.3.4. ERRoR InDICAtIon
The commonest error in the case of an insulation measurement is the presence of a voltage on the terminals. If it is greater than
50 V, the insulation measurement is not enabled. Eliminate the voltage and start the measurement over.
The measurement may be unstable, probably because of an overly capacitive load or an insulation fault. In this case, read the
measurement on the bargraph.