20
Before the measurement: to display the measurements already recorded.
During or after the measurement: to record it.
The direction of the arrow indicates whether you can make a reading (arrow pointing out) or a recording (arrow
pointing in).
The percentage indicates the quantity of memory already used.
6 %
EARTH 3P
02
/
12
/
2009 10
:
47
50
.
0
Ω
50
.
1 Hz
R
E
3 2 . 0 8
Ω
R s 1 . 5 8 k
Ω
R h 1 . 3 2 k
Ω
3 %
..\..
Measurement result.
Resistance of the S rod.
Resistance of the H rod.
Case where the measurement is
below the alarm threshold.
3.4.3. READInG of tHE REsUlt
In the case of a detailed measurement:
Press the TEST button to start the measurement. The measurement stops automatically.
This symbol invites you to wait while the measurement is in progress.
TEST
Value of the alarm threshold.
Compensation for the resistance of
the measurement leads is activated.
If the measurement must be made in a damp environment, remember to change the value of maximum contact voltage U
L
in
Setup (see §5) and set it to 25 V.
The key is used to see the voltages
before the beginning of the test.
Do not forget
to reconnect the earth strap
at the end of the measurement
before powering the installation back up.