8. R-SW handpiece Specifications and Operation
• Clean and disinfect the transmitter and the transmitter screw cap with the usual alcohol-based cleaning agents and disinfec-
tants.
• Dry the transmitter and transmitter screw cap before screwing them together.
• Push the insert into the front cap and screw the two cap parts together until finger-tight.
NOTE
Make sure that the two cap parts of the D20 and the D35 shock transmitters are screwed firmly in place and that the shock
transmitter screw cap is screwed firmly to the shaft.
• Check the screw connection of the shock transmitter screw cap and cap parts during prolonged treatment phases
8.5 R-SW HANDPIECE OVERHAUL
Shock waves are generated mechanically. Due to the effects of friction, the handpiece components are continuously exposed to
mechanical stress which will cause minor wear.
NOTE
The R-SW handpiece should be overhauled about every 1,000,000 shocks. This can be done quickly and easily by the user of the
instrument. All that is required is the overhaul kit that includes all required wearing parts.
8.5.1 CONTENTS OF THE R-SW OVERHAUL KIT
2 projectiles
2 guide tubes
2 socket sealing rings
1 O-ring guide
The overhaul kit is available from your dealer, part number 23825.
NOTE
The sealing rings, the projectile and the guide tube must always be replaced each time the handpiece is overhauled.
• Observe the O-ring Guide when selecting the sealing rings to be used. It is contained in the overhaul kit.
DANGER
If the unit is not safe for operation, it must be repaired by a certified service technician and the operators must be
informed of the dangers posed by the unit.