23
Mobile 2 RPW User Manual
4. Cleaning, Maintenance, Overhaul
4.4 MAINTENANCE
Preventive maintenance is not necessarily required. However, regular maintenance may help to identify possible defects at an
early stage and thus increase the safety and service life of the equipment. Maintenance services can be ordered from our regional
representatives in your area.
We recommend that functional and safety checks be performed at least once a year. National accident prevention regulations and
test and inspection intervals prescribed for medical devices must, of course, be observed.
NOTE
For further details on content and performance of the safety checks please contact your local dealer.
The following checks should be performed to ensure that the device operates safely.
1. Earth leakage current test in accordance with national regulations.
2. Earth impedance test (incl. handpiece housing and with mains cable) in accordance with national regulations.
4.5 DISPOSAL
When disposing of the present medical products, no special measures have to be observed. Please proceed in accordance with
applicable country-specific regulations. After expiration of its service life, dispose the device as waste electronic equipment.
4.6 REPAIR
Repair work on defective instruments must only be carried out by personnel suitably authorised by DJO LLC. Use only original parts
designated by STORZ MEDICAL AG and supplied by DJO LLC. The personnel suitably authorised can be from DJO LLC or be repre-
sentatives of DJO LLC agencies and dealers.
4.7 SERVICE LIFE
The average expected service life (MTTF) in accordance with EN60601-1:2005 + A1:2012 is
–
approximately 3 500 hours for the medical - electrical device Chattanooga Mobile 2 RPW USA.
For information about the service life of your handpiece, please refer to section 8.9 (Service life R-SW handpiece) and section 9.14
(Service life V-ACTOR handpiece).
Exceeding the service life can be expected to result in a failure of the instrument and accessories.
This also applies to handpieces.
No warranty claims shall be accepted beyond the information given in chapter 8.