9. V-ACTOR II Specifications and Operation
9.7 START-UP
• Connect the V-ACTOR II handpiece to the control device.
• Set the energy in V-ACTOR operating mode to an initial value of 2 bar.
• Activate the V-ACTOR II trigger button.
NOTE
The trigger button functions as an on/off switch when it is pressed briefly. Pressing it for longer causes it to function as a tip
switch, i.e. the pulses will continue until the button is released.
9.8 FUNCTIONAL CHECKS
• Set the energy level in V-ACTOR mode to 2.4 bar.
• Reset the actual number of pulses on the parameter display of the control panel.
• Release pulses with a frequency of 31 Hz.
• Check that the triggered pulses are correctly counted on the treatment pulse counter of the control device.
9.9 STANDARD SETTINGS
• Start the V-ACTOR II treatment at an energy level of 2 bar and a frequency of 20 Hz.
• For more detailed instructions, refer to the application brochure and the treatment recommendations.
9.10 TREATMENT
CAUTION
Read
Chapter 3.1 General Warnings And Safety Information
before beginning treatment.
NOTE
The maximum energy level used during treatment must not cause the patient undue pain under any circumstances.
• Apply a sufficient amount of massage oil to the patient’s skin in the treatment area and to the V-ACTOR II transmitter.
• Perform the V-ACTOR II treatment as recommended in the application brochure / treatment recommendations.
CAUTION
The handpiece may not be operated while idling (without an impact surface).
•
Do not trigger pulses unless the pulse transmitter is in contact with the treatment zone!
CAUTION
We recommend that the user and the patient wear suitable hearing protection.
•
Always offer the patient hearing protection.