12
Mobile 2 RPW User Manual
2. Installation Instructions
2.3.3 HANDPIECE CONNECTION
•
Connect the plug of the handpiece to the handpiece connection (Fig. 2 - 4/1) on the device.
Fig. 2 - 4 Handpiece connection
•
Make sure that the red dots on the connector match the red dots on the handpiece connection (Fig. 2 - 5).
Fig. 2 - 5 Connecting the handpiece
•
Place the handpiece into the handpiece holder.
2.3.4 POTENTIAL EQUALISATION (OPTIONAL)
The device features a potential equalisation connection (Fig. 1 - 3/5). Where necessary, connections for potential equalisation must
be made by suitably qualified personnel.
The potential equalization connection must be connected in accordance with the relevant national regulations.
1
CAUTION