47
Mobile 2 RPW User Manual
9. V-ACTOR II Specifications and Operation
• Gently pushing the plug into the connector will immediately lock the connection and will prevent the connector from disenga-
ging automatically when the cable is pulled.
• To break the connection, pull the outside of the plug body. This means the locking function is initially released, then the plug is
pulled out of the socket.
9.4 REPLACING THE TRANSMITTER
– V25
• To remove the 25 mm transmitter, unscrew the transmitter screw cap from the handpiece and pull the transmitter out.
• Clean all parts of the transmitter as described in
Chapter 9.11 Cleaning
.
• Reassemble the transmitter in reverse order.
• Screw the new transmitter onto the handpiece until finger-tight.
1
2
Fig. 9 - 4 Replacing the transmitter V25
– V40
• To remove the transmitter V40, unscrew the transmitter from the handpiece.
• Unscrew the transmitter screw cap and pull the transmitter out.
• Remove the sealing ring by pressing it apart at the cut-through point.
• To clean, press the spring element together slightly and remove the residues underneath it.
• Clean all parts of the transmitter as described in
Chapter 9.11 Cleaning
.
• Reassemble the transmitter in reverse order.
• Make sure that the smooth side of the sealing ring is in contact with the transmitter head.
• Screw the new transmitter onto the handpiece until finger-tight.
1
2 3 4
5
6
1
3
5
6
2 4
Fig. 9 - 5 Transmitter V40
1 Front cap
2 Transmitter head
3 Sealing ring
4 Spring element
5 Rear cap
6 Cut-through point