3. Operation
3.3 FUNCTIONAL CHECKS
Perform the following functional checks after the instrument has been installed:
•
Check the control device and handpiece for damage.
•
Put the device into operation.
•
Set the pressure to 1.6 bar.
•
Reset the treatment shock counter (Fig. 3 - 1/2) with the reset button (Fig. 3 - 1/4) on the front of the instrument.
•
Release individual shocks in single shock mode.
•
Release shocks in continuous shock mode (shock frequency 1 Hz and 15 Hz).
•
Check that the triggered shocks are correctly counted on the treatment shock counter on the front of the instrument.
•
Set the pressure to maximum 5 bar.
•
Release individual shocks in single shock mode.
•
Release shocks in continuous shock mode (shock frequency 1 Hz and 8 Hz).
•
Test the other frequencies as follows:
Operating Mode
Energy/Frequency
R-SW
4.0 bar / max. 21 Hz
R-SW
4.6 bar / max. 15 Hz
R-SW
5.0 bar / max. 12 Hz
V-ACTOR
3.0 bar / max. 31 Hz
V-ACTOR
3.8 bar / max. 22 Hz
V-ACTOR
5.0 bar / max. 13 Hz
3.4 STANDARD SETTINGS
•
Before each treatment, set the treatment shock counter (Fig. 3 - 1/2) on the control device to zero by pressing the reset
button (Fig. 3 - 1/4).
•
A total of about 2,000 shocks must generally be applied per therapy session. Please refer to the device application
brochure for details.
•
Start the treatment at a pressure of 1 bar and a frequency of 5 Hz.