15
Mobile 2 RPW User Manual
3. Operation
Fig. 3 - 1 Setting the energy on the control device
1 Display of selected shock frequency
2 Treatment shock counter
3 Display of selected pressure (nominal value)
4 Shock counter reset button
5 Dial for setting the pressure
6 Handpiece
7 Handpiece connector
8 Buttons for setting the shock frequency
The maximum application pressure is limited to 5 bar. To ensure correct system operation, a minimum pressure of 1.0 bar is required.
•
Press the trigger button on the handpiece.
•
To work in single shock mode, select the “-” symbol (dash) in the “Frequency” selection box and activate the trigger button.
•
To work in continuous shock mode, select a continuous shock frequency in the range from:
Operating Mode
Energy/Frequency
R-SW
1.0 - 4.0 bar / max. 21 Hz
R-SW
4.6 bar / max. 15 Hz
R-SW
5.0 bar / max. 12 Hz
V-ACTOR
1.4 - 3.0 bar / max. 31 Hz
V-ACTOR
3.8 bar / max. 22 Hz
V-ACTOR
5.0 bar / max. 13 Hz
•
Activate the trigger button.
NOTE
A pressure change automatically entails a change in frequency if the set frequency exceeds the maximum permitted
frequency (see Table 3 - 1)
3
2
1
5
4
8
6
7