5. Error Messages and Trouble-shooting
5.2 TROUBLE-SHOOTING
CAUTION
Unplug the mains cable from the instrument before you carry out any maintenance work!
Fault description
Possible cause
Corrective action
Instrument does not work
•
Power failure
•
Defective mains fuse
•
Defective mains plug
•
Check the power supply
•
Replace the fuses
•
Replace the mains cable
No compressed air supply
•
Leaks in handpiece cable or cable not
properly connected
•
Clogged compressor air filter
•
Check the cable and tube connections
and replace them, if necessary
•
Check the compressor air filter and
replace it, if necessary
No shock wave power output
•
No compressed air supply
•
Blocked or worn projectile
•
Malfunction in control device
•
Handpiece defective
•
Dismantle the handpiece
•
Clean the guide tube and projectile
•
Overhaul the handpiece
•
Call your Service centre
•
Replace the handpiece