Chapter 3
Operation
52
SAFETY FIRS T
Edition 1 – December 2019
MSA G1 Classic Breathing Apparatus
Step
Description
Photos
5.4
Do not move the DSU.
Observe that it goes through the first alarm and
into the second alarm (after approximately
40 seconds).
5.5
Cancel the second alarm.
Press the MANUAL ALARM button twice
within one second.
5.6
Activate the MANUAL ALARM, by depressing
the centre button.
5.7
Turn off the DSU.
Hold the ON/OFF button and the ALARM
button concurrently for a minimum of four
seconds.