5. Menus
5.2 Window menu
62
Set the detection type, including dynamic refresh, minimum value, maximum value, and average four
trace types, and the default is dynamic refresh.
[Filter Selectivity]
Set the filter type, including auto, regular, narrow, and sharp four selection types, and the default is
regular.
[Measurement Mode]
Set the measurement mode, including two modes: periodical and continuous, and the default is
periodical mode.
[Measurement Time Mode]
Set the type of measurement time, including automatic and manual modes. Auto mode is selected by
default (in the auto mode the measurement time is automatically calculated by the receiver).
[Measurement Time]
Set the measurement time, the range is 500μs-900s. The default value is set to 500μs and valid when
the measurement time mode is manual.
[Frequency Step Mode]
Set the frequency step mode including the automatic and manual modes, and the automatic mode is
selected by default.
[Frequency Step Interval]
Set the frequency step interval which is valid when the frequency step mode is manual. The range is
1Hz~1GHz and the default value is set to 1MHz.
[Differential Mode]
Set the differential display of the spectrum. Differential mode will display when it is on, and it is in normal
display when it is off.
2) Display
The display-related property pages under the IF Panorama window are shown in Fig. 5.41.
Fig. 5.41 The display-related property pages under the IF Panorama window
The parameters for the display-related property pages are described below:
[Highest Level]