3 Quick Start
3.4 Data Management
28
3.4.2.1 Data file type
The types of data files related to file management are shown in Table 3.8.
Table 3.8 Data File Types
File type
Description of data storage
Startup status file
Store the measurement status of the current
instrument, and call it at startup.
Reset status file
Store the measurement status of the current
instrument for the user to call during resetting
Saving of trace file by user
Store the trace data of the current trace
Saving of image file by user
Store the bitmap of the current operation screen in png
format
Saving of audio data by user
Store the audio file saved by users
Saving of IQ data file by user
Store the I/Q modulation data
Antenna factor file
Store the antenna factor information
Storage table file
Store the storage table information edited by the user
Suppression table file
Store the suppression table information edited by the
user
Map file
Store map data
3.4.2.2 File Saving/Loading Method
The 3943B monitoring receiver provides data file saving/calling function. File saving is to store the
measurement data in a file according to the agreed format (for example: *. png, *. dat). File call refers to
opening the selected data file and refreshing display information, such as measurement parameters and
list parameters, so as to facilitate the observation and evaluation by users.
According to different information types and management methods, the 3943B monitoring receiver
provides seven kinds of file saving functions and six kinds of file calling functions. The seven types of
files that can be saved are: reset status file, trace file, image, audio, storage table, suppression table and
IQ data file. The file types that can be called are: reset status file, trace file, storage table, suppression
table, antenna factor and audio file. Users only need to enter the corresponding menu to pop up a dialog
box, and specify the file name to be saved or select the corresponding file, so as to save and call the file.
Specific instructions are as follows:
For the saving and calling of reset status file, refer to 3.4.1 ―Reset status storage/ calling‖;
Saving and calling of trace, audio and IQ data files are introduced in the recording and playback option
functions.
See 5.1.3 ―List menu‖ for the saving and calling of storage table and suppression table files;
See 3.4.3 ―Screen shot storage‖ for the saving and calling of image files.
3.4.2.3 File directory management
The 3943B monitoring receiver provides file directory management function. See 5.1.5 ―File menu‖ for
file browsing, copying, cutting, pasting, deleting and other operations.
3.4.2.4 File Format Description
All files are directly stored in the monitoring receiver. Except IQ data and antenna factor files, other files
are in binary file format, and the file suffix is determined by the specific file type.
3.4.3 Screen shot storage
The 3943B monitoring receiver provides the function of storing screen shots in graphic files (png format).
Press [Screen Shot] menu on the main interface to pop up the save screen dialog box, as shown in Fig.
3.22.