SC310
User’s Manual
61
5.9 Measurement range (octave band spectrum analyser)
L
T
Function
for C-130 + PA-13:
Measurement range (with linearity error lower than 0.4 dB):
Octave bands with nominal central frequency
Lower than 63 Hz:
Upper limit:
Lower
limit:
Higher than 63 Hz:
Upper limit:
Lower
limit:
The noise (elec thermic of the microphone) is, as a minimum, 10 dB
lower than the lower limit of the measurement range.
137
dB
25
dB
137
dB
20
dB
for C-250 + PA-14:
Measurement range (with linearity error lower than 0.4 dB):
Octave bands with nominal central frequency
Lower than 8 kHz:
Upper limit:
Lower
limit:
Higher than 8 kHz:
Upper limit:
Lower
limit:
The noise (elec thermic of the microphone) is, as a minimum, 10 dB
lower than the lower limit of the measurement range.
137
dB
18
dB
137
dB
20 dB
5.10 Third octave band filters
Frequency evaluation system
Base 10
Reference attenuation
0 dB
Operative linearity range
same as the measurement range
Third octave band nominal central frequencies
Nominal central
frequency
Exact base 10
frequency
Nominal central
frequency
Exact base 10
frequency
20 Hz
19.95 Hz
500 Hz
501.19 Hz
25 Hz
25.12 Hz
630 Hz
630.96 Hz
31.5 Hz
31.62 Hz
800 Hz
794.33 Hz
40 Hz
39.81 Hz
1000 Hz
1000.00 Hz
50 Hz
50.12 Hz
1250 Hz
1258.93 Hz