SC310
User’s Manual
69
Set of accessories tested in the verification of
the electromagnetic compatibility requirements.
The ones described in chapter
5.29.
Configuration for the normal mode of working.
Sound level meter mode, L
AF
function.
Configuration for the reference orientation.
The SC310 in vertical position,
with the SC310 main axis
(preamplifier) perpendicular to
the field propagation direction
It is not observed any emission difference from the normal mode of operation being the
equipment in vertical position, with all cables connected and with the AM240 mains
feeder.
The equipment shows no degradation or loss of function after being exposed to
electrostatic discharges.
5.21 Vibration influence
For frequencies from 20 to 1,000 Hz and 1 m/s²:
< 75 dB(A)
5.22 Batteries & External supply
Battery
Two AA sized (LR6) 1.5 V batteries
Typical battery life with continuous use:
Sound
Level
Meter
Spectrum Analyser 1/1
Spectrum Analyser 1/3
The typical battery life with the display lighted can be reduced a 50% respect
the values above
15:00
hours
13:00
hours
11:30
hours
External Supply
Voltage input range
3.8 to 4.2 V DC
Minimum supply current
100 mA
For DC supply use the
AM240 or AM241model mains feeder
does not accept any responsibility in the case of the use of t
main feeders other than the recommended.
5.23 Dimensions and weight
Dimensions:
341 x 82 x 19 mm
Weight:
With battery 550 g
Without 500
g