Replacing the AC Fuses
A fuse may need to be replaced if the Oximeter is plugged into an electrical outlet but the
AC Power indicator is not illuminated. The Oximeter has a dual-fuse AC power input receptacle
(both AC lines are fused).
Warnings:
Before changing a fuse, turn the Oximeter off, unplug the AC power cord, and
remove the Battery as described in
Removing the Battery on page 65
.
For continued protection against fire hazard, replace a fuse only with an
identically rated fuse. See
Specifications
on page 89
Other fuses located inside the Oximeter and Battery are not user-replaceable,
and must be replaced only by CASMED-authorized Service personnel.
To replace the AC power fuses:
1.
Turn the Oximeter off, disconnect the AC power cord, and
remove the Battery as described on page 65.
2.
Insert a small, flat screwdriver under the edge of the AC fuse
cover (indicated by the arrow in Figure 33) and pop the cover
open.
3.
Use the screwdriver to gently pry up the red fuse holder from
within the AC Fuse compartment and remove it.
4.
Remove the suspect fuse.
5.
Place a new fuse into the holder as indicated in Figure 34.
6.
Repeat this process for the second fuse.
Figure 34:
AC Fuse Placement
7.
Insert the fuse holder back into the AC Fuse compartment (the fuse holder can be inserted
in either orientation).
8.
Close the AC Fuse compartment and press down until you hear an audible click.
9.
Reconnect the AC power cord to the Oximeter.
Figure 33:
AC Fuse Cover
Correct Placement
Incorrect Placement
PN 21-22-3000 Rev 07
67
FORE-SIGHT ELITE User Manual