Physical Placement
Warnings:
The Oximeter needs special precautions if it is placed close to a strong
EMC transmitter such as an X-ray or ESU device. There is a risk that these
signals could interfere with the Oximeter and disrupt its performance, causing
measurements to be inaccurate during use of such equipment.
To reduce the risk of injury, do not move or place the Oximeter or accessory in
any position that might cause it to fall on a patient, bystander, or operator.
Do not lift or pull the Oximeter by any cable. As with all medical equipment,
carefully route and secure all cables to reduce the risk of patient entanglement
or strangulation.
Avoid placing the Oximeter where a patient could access the controls.
Caution:
Avoid placing either side of the Oximeter against objects that could restrict air
flow through the air intake vents (left side) and exhaust vents (right side).
Mounting Solutions
An optional mounting plate (P/N 01-06-0239) is available for joining the Oximeter with CASMED-
approved mounting solutions; the four non-skid feet may also be removed to facilitate installation.
Contact your CASMED distributor for more information.
Warning:
The proper assembly of the various mounting solutions cannot be assured
for structural integrity. There is risk that a damaged or improperly assembled
Oximeter mounting solution may fall on patient, bystander, or operator.