Configuring StO
2
Alarm Limits
StO
2
limits (both low and high) can be set independently for each channel. The default values
are 50% (low) and 90% (high), and any StO
2
reading outside that range will trigger an alarm.
The values can be adjusted if necessary.
Cautions:
Setting the Low Alarm limit to an extremely low value can render the Low Limit
detection ineffective.
Likewise, setting the High Alarm limit to an extremely high value can render the
High Limit detection ineffective.
To change an StO
2
alarm limit:
1.
Access the Patient Setup screen by touching a numeric or location icon in the Channel
Status area.
You can also, from the Main Menu, touch
Patient
and then touch
Patient Setup
.
2.
Touch the button for the channel limit you wish to modify.
3.
Use the keypad to enter the new value.
Allowable ranges are 3-93 for Low and 8-98 for High.
Note:
For a given channel, the High cannot be set lower than the Low or vice-versa.
There must be a minimum 5% difference between the Low and High limits.
4.
Touch
✔
to save your changes and close the Patient Setup screen, or touch
✘
to cancel.
PN 21-22-3000 Rev 07
45
FORE-SIGHT ELITE User Manual