8) Error Message
Error 01
▣
Reason
The weight is unstable to initialize the scale
Solution
☞
Put the scale on a stable place and turn on the power
Error 02
▣
Reason
Load cell connection failure or error in A/D conversion part.
Solution
☞
Check the load cell connector to see if the polarity of signal is reversed
Error 05
▣
Reason
You pressed any key for long time or problem of key part
Solution
☞
If there is no problem in key part, call your CAS dealer
Error 08
▣
Reason
You have set the operating condition of ZERO key or TARE key not to
operate when the indicator is not stable.
Solution
☞
Reset the operating condition of ZERO and TARE key in SET mode F09
Error 09
▣
Reason
Current weight deviates from zero range
Solution
☞
Set the operational range of the ZERO key within 2% or 10% of the
maximum capacity in SET mode F08
Error 10
▣
Reason
Tare weight exceeds the maximum capacity of the scale
Solution
☞
Tare weight should be lower than the maximum capacity.
Otherwise, the maximum capacity should be greater than the tare
Error 13
▣
Reason
The zero range deviates from the set range
Solution
☞
Confirm that there is nothing on the platter.
If nothing exists, do calibration in CAL mode
Summary of Contents for CI-6000A Series
Page 1: ...2 ...
Page 8: ...9 4 Measure of Appearance ...
Page 22: ...23 5 Sealing Method Sealing method of cal switch Sealing method of Load Cell Connector ...
Page 51: ...52 Normal Batching Diagram Loss in Weight Batching Diagram ...
Page 52: ...53 USER PROGRAM CONTROLMODE Normal Batching ...
Page 54: ...55 USER PROGRAM CONTROLMODE Loss in Weight ...
Page 56: ...57 AUTOMATIC PROGRAM CONTROLMODE Normal Batching ...
Page 58: ...59 AUTOMATIC PROGRAM CONTROLMODE Loss in Weight ...
Page 71: ...72 MEMO ...
Page 72: ...73 MEMO ...
Page 73: ...74 MEMO ...
Page 74: ...75 ...