44
TEST 2
FUNCTION : Display test
KEY
DISPLAY
DESCRIPTION
ENTER: Go to menu
Selection mode
tESt 2
TEST 2 condition
Other keys : Perform test
8888888
TEST 2 is performed
TEST 3
FUNCTION : A/D Conversion test
KEY
DISPLAY
DESCRIPTION
tESt 3
TEST 3 condition
: Change gain.
: Change filter
ENTER: Go to menu
Selection mode
97853
Shows digital value of current weight.
This value means converted digital value.
Note 1. Check whether digital value is changing. If the digital value is fixed or zero is
displayed, please check the connection of load cell.
TEST 4
FUNCTION : RS-232C test with computer (COM1)
KEY
DISPLAY
DESCRIPTION
tESt 4
TEST 4 condition
00------00
Wait for transmission and reception
03------00
Transmit : 3, Receive : none
: Transmit to PC
after increasing
value.
: Transmit to PC
after increasing
value
ENTER: Go to menu
Selection mode
08------49
Transmit : 8, Receive : 1
Note 1. Do this test in Hyper Terminal of PC after the connecting serial port with PC.
- Port Selection in Hyper Terminal has the Flow Control item, you have to set nothing.
Note 2. Send no.1 in computer keyboard and check if indicator receives no.49
Send no.8 in indicator key( ) and check if computer receives no.8
Note 3. Baud rate should be specified in SET mode before testing.(F31)
Summary of Contents for CI-6000A Series
Page 1: ...2 ...
Page 8: ...9 4 Measure of Appearance ...
Page 22: ...23 5 Sealing Method Sealing method of cal switch Sealing method of Load Cell Connector ...
Page 51: ...52 Normal Batching Diagram Loss in Weight Batching Diagram ...
Page 52: ...53 USER PROGRAM CONTROLMODE Normal Batching ...
Page 54: ...55 USER PROGRAM CONTROLMODE Loss in Weight ...
Page 56: ...57 AUTOMATIC PROGRAM CONTROLMODE Normal Batching ...
Page 58: ...59 AUTOMATIC PROGRAM CONTROLMODE Loss in Weight ...
Page 71: ...72 MEMO ...
Page 72: ...73 MEMO ...
Page 73: ...74 MEMO ...
Page 74: ...75 ...