92
(
Selecting one of standard Settings)
1
From the mass display, press the
[CAL]
key
repeatedly until the
[intFACE]
display appears.
Press the
[O/T]
key. The display shows “iF;F1”
2
If necessary, press the
[CAL]
key repeatedly
until the desired standard setting display appears.
Then, press the
[O/T]
key.
3
Press the
[POWER]
key repeatedly to return to
the mass display.
14.3.3 User Setting
The user setting allows individual setting for each item in communication settings.
(
Making user settings)
1
In mass display, press the
[CAL]
key repeatedly
until “intFACE” appears and press the
[O/T]
key. “iF:F1” appears.
2
Press the
[CAL]
key repeatedly until “iF: USEr”
appears and press the
[O/T]
key. “io.b:* ***”
appears (communication speed settings).
(**** shows the current setting; 2 to 4 characters,
the same applies hereafter.) After this, pressing
the
[CAL]
key alternates the display in this order:
“io.d: ****” (delimiter setting), “io.P; *:**”
(parity setting), “io.S: ****” (stop bit setting),
“io.F: ****” (data format setting), and
“io.H: ****” (handshake setting).
3
When the item to be set appears, press the
[O/T]
key. (Refer to the following selections thereafter.)
(When F1 is selected)
(When F1 is selected)
(Example)
(Example)
(Example)
(Example)
(Example)
(Example)
Displays of all the items